TRN-QRG-100125-US(AC)
Initiating Manual Mode
NOTE:
Manual Mode must be enabled in a settings group before it can be utilized.
Please refer to Operation Manual for instructions to enable access and activate Manual Mode in a new Settings Group.
This feature enables the user to control the Cell Saver Elite+ device manually for cell salvage with minimal sensor monitoring.
To control the device, the Fill, Wash, Empty, Concentrate, and Return phases are manually selected by the user.
In Manual Mode:
The line sensor and bowl optics are disabled
The air detector is used for volume accounting in the Fill, Wash, and Concentrate phases,
and to determine when to end the Empty and Return phases
To initiate Manual Mode:
1. After starting a procedure using a settings group that
has Manual Mode enabled, touch Active Settings.
2. Select
Manual Mode
from the drop-down list.
The
Switching to Manual Mode
screen appears.
3. Touch
Confirm
to switch to Manual Mode.
- The device remains in the same phase it was in prior
to the transition to Manual Mode (for example, Fill,
Wash, and so on)
- All processing phases are now available from the
touch screen
4. The device is now in
Manual Mode
.
To exit Manual Mode:
1. Touch
Active Settings
to open the drop-down list.
2. Select either the same settings group that was being
utilized immediately prior to switching to
Manual Mode
or a different settings group.
NOTE:
Upon exiting Manual Mode, if the bowl is not empty and the
device is in the Fill or Concentrate phase, the device returns
the contents of the bowl to the reservoir; if the bowl is not
empty and the device is in the Wash, Return, or Empty phase,
the device continues with that phase.
This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.