background image

For a list of worldwide office locations and contact information, visit www.haemonetics.com/officelocations

 

 
© 2020, 2022 Haemonetics Corporation. Haemonetics, Cell Saver, Elite, and SmartSuction are trademarks or registered trademarks  
of Haemonetics Corporation in the USA, other countries, or both. All other product names, trademarks and registered trademarks are the  
property of their respective owners. 11.2022 USA. TRN-QRG-100125-US(AC)

Summary of Contents for Cell Saver Elite+

Page 1: ...This guide is not intended as a substitute for the Cell Saver Elite System Operation Manual or authorized training Quick Reference Guide ...

Page 2: ...ce and allow system self test to finish 5 Select bowl size or scan processing set when prompted Load reservoir and A A line 1 Place reservoir in weigher and close drain port slide clamp 2 Connect vacuum line to device vacuum port and yellow capped port on reservoir If using external suction connect external vacuum to yellow capped port on reservoir 3 Receive and attach A A line to blue capped port...

Page 3: ...s are open and twist lock connection is secure 3 Thread pump tubing around pump and place tubing manifold into position 4 Floss tubing into air detector and insert clear tubing and color coded lines into grooves in valve module 5 Close pump platen and close and latch valve module cover 6 Place bowl firmly into centrifuge with lower inlet port facing left NOTE If using 70 mL bowl first place white ...

Page 4: ...ervoir and open drain port slide clamp 10 Hang wash solution on lower pigtail of right IV pole 11 Close both yellow line tubing clamp s and spike saline bag s 12 Open yellow line clamp s on saline bag s to be used Initiate procedure 1 Close device cover 2 Touch Active Settings Haemonetics Default or Fat reduction 3 Touch Start Procedure NOTE System will begin processing when minimum fluid level is...

Page 5: ... bag of sterile 0 9 saline for injection USP Anticoagulant ACD CPD or 25 000 30 000 IU Heparin 1 liter bag sterile 0 9 saline for injection USP Cell Saver Elite processing set Surgical collection reservoir Aspiration and anticoagulant A A line Filtered vacuum line Vacuum source Regulated to 150 to 200 mmHg for Intra op use or use integrated SmartSuction Centrifuge chuck adaptor Reusable for 70 mL ...

Page 6: ...rvoir 3 Open Post op suction set receive and attach Y connector to wound drain tubing 4 Disconnect A A line and discard according to the facility s policies and procedures for handling of biohazardous materials 5 Connect Post op suction set to reservoir 6 Power down device disconnect from power lower IV poles and transport device to PACU NOTE Do not select End Procedure 7 Connect device to power p...

Page 7: ...me precautions Emptying the waste bag Drain waste fluid into empty container such as irrigation solution bottles for discard NOTE Unless the bowl is completely empty keep fluid level in the waste bag above the 1 liter mark on the waste bag This prevents loss of displacement air Changing the waste bag To prevent loss of displacement air the waste bag can be changed ONLY at the end of the EMPTY cycl...

Page 8: ...led touch Active Settings 2 Select Manual Mode from the drop down list The Switching to Manual Mode screen appears 3 Touch Confirm to switch to Manual Mode The device remains in the same phase it was in prior to the transition to Manual Mode for example Fill Wash and so on All processing phases are now available from the touch screen 4 The device is now in Manual Mode To exit Manual Mode 1 Touch A...

Page 9: ...nue washing until the effluent line is clear NOTE If during the Wash phase the RBC supernatant interface disappears and RBCs begin to spill into the waste bag touch Pause to allow the interface to reappear then slowly increase to a rate that just maintains this interface usually 200mL min or less Emptying the Bowl 1 After the red cells have been processed touch Empty to send them to the reinfusion...

Page 10: ...lable from a Standby or Stopped state or when in Manual Mode or Quick Transfer Mode Quick Transfer Mode When time is of the essence for returning the product to the patient due to trauma sudden massive blood loss thoracoabdominal aneurysm and so on users can initiate Quick Transfer This mode quickly moves the shed blood from the reservoir to the reinfusion bag The device automatically cycles throu...

Page 11: ...ts to the previous settings group and enters Standby NOTE Emergency Mode is not available with the 70 mL bowl NOTE The line sensor is not active during Emergency Mode To enter Emergency Mode 1 Touch Emergency Mode 2 Touch On to confirm Pump speed increases to process faster RBCs may spill into waste bag during Emergency Mode Device will continue FILL WASH EMPTY until no more RBCs remain in reservo...

Page 12: ...are open Ensure the slide clamp on the outlet port of the Reservoir is open Check that the vacuum applied to the reservoir is 250 mmHg 33 kPa Dilution of packed red cells with saline is necessary to decrease viscosity within the Reservoir before initiating Quick Transfer To avoid clotting in the Reservoir ensure the volume of blood is properly anticoagulated before initiating Quick Transfer Initia...

Page 13: ... bowl of blood at the end of a Cell Saver Elite procedure or whenever it is necessary to process blood before a full bowl has been collected 1 Touch Wash to begin partial bowl wash An event message appears 2 You will have 3 options concentrate double wash volume or Single wash NOTE Blood processed using a partial bowl may have a lower hematocrit than blood processed using a normal full bowl ...

Page 14: ... Procedure Only one of these options can be selected to manage the last bowl Last Partial Bowl Wash Double The device transitions to the Last Partial Bowl Wash phase and automatically doubles the wash volume Single The device transitions to the Last Partial Bowl Wash phase using the normal wash volume Off The device omits the Last Partial Bowl Wash defaults to End Procedure option End Procedure On...

Page 15: ...e screen and follow screen instructions 3 Close all clamps on all disposable tubing 4 Cap all ports on collection reservoir NOTE Tubing stubs and caps included with reservoir and processing set may be used for closing open ports 5 Remove waste bag from support pins may be emptied before removing 6 Remove bowl from centrifuge well 7 Remove remaining tubing harness 8 Remove collection reservoir keep...

Page 16: ...ine to device vacuum port and yellow capped port on reservoir If using external suction connect external vacuum to yellow capped port on reservoir 3 Receive and attach A A line to a blue capped port on reservoir 4 Turn on suction If using manual suction set to minimal acceptable level 200 mmHg 5 Close roller clamp on A A line 6 Hang and spike anticoagulant bag 7 Prime reservoir until breakthrough ...

Page 17: ...ouch Active Settings 2 Choose the setting from the drop down list Edit settings group 1 Touch settings group to edit 2 Touch view 3 Use arrows to select a parameter 4 Use arrows to change parameter 5 Touch done to save parameters NOTE Haemonetics Default and Fat reduction are settings groups that cannot be edited or deleted Return to Cell Salvage screen 1 Touch done Create new settings group 1 Sel...

Page 18: ... Bacitracin Neomycin Polymyxin Betadine Hypertonic solution 3 NaCl 7 NaCl Dextrose solutions Hypotonic solution Sterile water Glycine Lactated Ringers in presence of Citrate Anticoagulant The following are trademarks or registered trademarks of the company listed in parentheses in the United States other countries or both Avitene C R Bard Inc Betadine Purdue Products Inc Gelfoam Pharmacia Upjohn C...

Page 19: ...on check suction wand for possible occlusion 4 Check junction of A A line and reservoir for occlusion 5 Ensure reservoir is not full 6 Try briefly increasing suction to clear line 7 If problem persists use alternate suction source Touch screen problems The screen is not refreshing properly and parts of the screen appear to be frozen 1 Touch Menu 2 Select an option from the drop down list that is d...

Page 20: ... sterile air 2 Check the tubing for correct placement in the air detector 3 Check the effluent tubing for correct placement in the effluent line sensor 4 Check the effluent tubing for kinks and occlusions 5 Check the blue and red tubing for kinks and occlusions 6 Touch Continue to continue This guide is not intended as a substitute for the Cell Saver Elite System Operation Manual or authorized tra...

Page 21: ...pt to save change Changing processing sets during a procedure 1 Touch End Procedure 2 Remove current processing set 3 Install new processing set 4 Touch Resume Procedure NOTE All procedure statistics from the procedure will be retained and suction can remain on throughout this process Changing bowl size during a procedure If an error message appears Bowl Size Mismatch 1 Check the bowl for proper i...

Page 22: ...ocedures select history then select the desired procedure and press view You can access procedures events and device records Consumables Includes list number lot number and expiration date for disposables and solutions Procedure Includes procedure volumes times surgery type surgeon operator ID and patient ID Device Includes software versions and device serial number 3 Touch Edit to edit the record...

Page 23: ...ide is not intended as a substitute for the Cell Saver Elite System Operation Manual or authorized training TRN QRG 100125 US AC Customer Support Request product information Place an order Track a shipment Return products Report a problem Request clinical support Ask a question or issue a complaint To contact Customer Service Phone 800 537 2802 Opt 3 Fax 800 860 1512 Email CustomerServiceNA haemon...

Page 24: ...2022 Haemonetics Corporation Haemonetics Cell Saver Elite and SmartSuction are trademarks or registered trademarks of Haemonetics Corporation in the USA other countries or both All other product names trademarks and registered trademarks are the property of their respective owners 11 2022 USA TRN QRG 100125 US AC ...

Reviews: