This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.
TRN-QRG-100125-US(AC)
Final Cycle
This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.
TRN-QRG-100125-US(AC)
Final Cycle
When Final Cycle is activated, the device continues processing until air is detected in Fill, after which the device automatically
completes the procedure according to the selected Final Cycle settings (Double Wash / Single Wash / End Procedure)
The final cycle options are Last Partial Bowl Wash or End Procedure. Only one of these options can be selected to
manage the last bowl.
Last Partial Bowl Wash
Double:
The device transitions to the Last Partial Bowl Wash phase and automatically doubles the wash volume
Single:
The device transitions to the Last Partial Bowl Wash phase using the normal wash volume
Off:
The device omits the Last Partial Bowl Wash (defaults to End Procedure option)
End Procedure
On:
Once air is detected, contents of the bowl will be returned to the Reservoir, the Blue Line will be purged and
End Procedure
will appear on screen. The only option is to end the procedure immediately.
To activate
Final Cycle
1. Select the Settings mode from the drop down menu
2. Select View to view the Settings Group
3. Select Final Cycle and touch
(Up)
or
(Down)
to change the value for that setting