This guide is not intended as a substitute for the Cell Saver
®
Elite
®
+ System Operation Manual or authorized training.
TRN-QRG-100125-US(AC)
Troubleshooting/Tips (page 1 of 3)
Troubleshooting scenarios
Problem
Possible cause
Action
Decreased air
flow / aspiration
problems
Air leak
1. Ensure vacuum line is securely connected to device and reservoir and
check for leaks.
2. Check that reservoir has no open tubes or ports.
3. Ensure A&A line has been correctly connected.
4. Check reservoir for leaks.
5. If problem persists, use alternate suction source.
Obstruction
1. Check vacuum tubing for obstruction or kinking.
2. Check A&A line for obstruction or kinking.
3. Have surgeon check suction wand for possible occlusion.
4. Check junction of A&A line and reservoir for occlusion.
5. Ensure reservoir is not full.
6. Try briefly increasing suction to clear line.
7. If problem persists, use alternate suction source.
Touch screen
problems
The screen is not
refreshing properly
and parts of the
screen appear to
be frozen.
1. Touch
Menu
.
2. Select an option from the drop-down list that is different from the one
currently displayed. The screen should refresh properly to display new screen.
3. Touch
Menu
.
4. Select previous option from drop-down list. The screen should refresh
properly to display the first screen.
The screen has
locked up and is
not responding
when touched.
1. Restart the device.
2. Upon restarting, the operator is presented with the option to continue with the
previous procedure or to start a new procedure. Select the desired option.