Troubleshooting
Handpiece vibrates, is very noisy
and/ or becomes warm in the
front part
Use of bent burs/ tools
Use a
different bur/ tool to test this. The
bur/tool itself can be tested with the
hadewe Bur Shaft Tester (Art. No.
(4990) to see if it is bent).
Maximum allowed speed of the tool
was exceeded, which caused a
stronger vibration
pay attention to
the tool (i.g. bur, capper and
instrument) manufacturers handling
details.
Bur/tool can be inserted only with
difficulty, or not at all.
Use of bent burs/ tools
Use a
different bur/ tool to test this. The
bur/tool itself can be tested with the
hadewe Bur Shaft Tester (Art. No.
4990) to see, if it is bent).
Chuck is dirty
Clean the chuck
according to the instructions in the
manual.
Bur does not clamp, or does not spin
Chuck is not tightened well
Tighten the chuck firmly according
to the instruction in the manual.
Knob of the handpiece can not be
moved easily
Chuck is dirty
Clean the chuck
according to the instructions in the
manual.
Unit does not work at all.
Pull out the mains plug to reset the
device
The unit shows different sound
levels at different sites of use
The sound level varies according to
the surroundings of the unit. If you
set up the unit directly in front of a
wall, the sound is reflected stronger,
as if it would, while standing further
away from a wall.
Handpiece motor stutters shortly
(appr. 1 sec) when turning unit on
In very rare cases this effect did
appear. It will not harm or damage
the device.