13 

Troubleshooting 

Handpiece  vibrates,  is  very  noisy 

and/  or  becomes  warm  in  the 
front part  

Use  of  bent  burs/  tools 

  Use  a 

different  bur/  tool  to  test  this.  The 
bur/tool  itself  can be  tested  with  the 
hadewe  Bur  Shaft  Tester  (Art.  No. 
(4990) to see if it is bent). 

 
 

Maximum  allowed  speed  of  the  tool 
was  exceeded,  which  caused  a 
stronger vibration 

 pay attention to 

the  tool  (i.g.  bur,  capper  and 
instrument)  manufacturers  handling 
details. 

 
Bur/tool  can  be  inserted  only  with 

difficulty, or not at all.

 

Use  of  bent  burs/  tools 

  Use  a 

different  bur/  tool  to  test  this.  The 
bur/tool  itself  can be  tested  with  the 
hadewe  Bur  Shaft  Tester  (Art.  No. 
4990) to see, if it is bent). 

 

Chuck  is  dirty 

  Clean  the  chuck 

according  to  the  instructions  in  the 
manual. 

 

Bur does not clamp, or does not spin 

Chuck  is  not  tightened  well 

 

Tighten  the  chuck  firmly  according 
to the instruction in the manual. 

 

Knob  of  the  handpiece  can  not  be 

moved easily 

 

Chuck  is  dirty 

  Clean  the  chuck 

according  to  the  instructions  in  the 
manual. 

 
 

Unit does not work at all. 

 
 

Pull  out  the  mains  plug  to  reset  the 
device 

 

The  unit  shows  different  sound 

levels at different sites of use 

 

The  sound  level  varies  according  to 
the  surroundings  of  the  unit.  If  you 
set  up  the  unit  directly  in  front  of  a 
wall, the sound is reflected stronger, 
as  if  it  would,  while  standing  further 
away from a wall. 

 

Handpiece  motor  stutters  shortly 

(appr. 1 sec) when turning unit on 

 

In  very  rare  cases  this  effect  did 
appear.  It  will  not  harm  or  damage 
the device. 

 
 

Summary of Contents for 0897

Page 1: ...Manual Xantos Art No 0897 0898 FB 04 280 Rev 1 Art no 0897 0898 ...

Page 2: ...ctions 3 Cleaning Disinfection 7 Getting to know the unit 8 Before the use 9 Area of application 9 Installation site location 9 Maintenance Care 11 Cleaning the chuck 11 Handling details 12 Troubleshooting 13 Technical data 14 ...

Page 3: ... extracted air can escape well Only switch on the unit if the handpiece rests secured in the handpiece holder or if you hold it safely in your hand To avoid the risk of electrocution only connect the device to the mains power supply If you pull the power plug it will stop the connection to the mains power supply Set up the unit so that the power plug is always accessible The power switch should al...

Page 4: ...ction or spray technique Wear a nose and mouth protection Always follow the safety instructions provided by the manufacturer of the materials being formed and used For safety reasons wear goggles during the treatment Ensure that the patient s foot and the handpiece are held firmly in place during the operation To prevent injury work with extreme caution on the foot of the patient and use only ligh...

Page 5: ...e Before using the equipment assure that the tool can be pushed far enough into the chuck Test this with a dipstick and clean the handpiece if necessary the inserted tool sits properly in the chuck and will not cant jam while clamping it the tool is approved for using it at the max speed see manufacturer s instructions or only use the speed required by the instrument the tool is absolutely dry For...

Page 6: ...nly to be accomplished by an authorized specialist The mains cable may also only be changed by the manufacturer or an authorized specialist Never remove the housing especially not while the equipment is plugged into the socket since some components are under electric voltage Any unauthorized or improper handling e g trying to repair the unit of the equipment will immediately nullify the guarantee ...

Page 7: ... capper etc first Then handle the cleaning of the tool according to the instructions of it s manufacturer Remove tissue remains by wiping off the unit with a mildly moistened paper cloth Afterwards clean all surfaces with disinfection cloth e g Mikrozid AF Tücher of Schülke and Mayr The charging unit is only to be cleaned with a dry cloth ...

Page 8: ...8 Getting to know the unit 1 Speed display 2 LED shines during left rotation 3 Increase speed 4 change directory 5 decrease speed 6 on off button 8 ...

Page 9: ...Dental technique laboratory Milling trimming and polishing work at plaster casts and models Do not use this unit for dentist purposes i e working directly on the patient Industry Grinding milling and polishing works Installation site location The installation of the workroom must correspond to any regulations according to your country laws Set up the control unit so that the ventilation slits are ...

Page 10: ...ith a diameter of 2 35 mm this is the diameter usually used for tools in podiatry To open the chuck push the spanner forward with your thumb and leave it there Now insert the tool as far as possible Pull back the knob and the tool will be firmly clamped For security reasons only change the tool if the handpiece is turned off Setting up the speed Hold the handpiece incl tool in your hand or store i...

Page 11: ...k To assemble the chuck follow the steps in reverse order Make sure that the chuck is opened during the assembly the handpiece button must be in forward position and is firmly applied afterwards Cleaning the chuck Soak the chuck in spirit or a special cleaning liquid for tools Clean the slits and the hole of the chuck with a dry scrubber or brush Rinse the chuck very well and make sure you let it ...

Page 12: ...when you not working with the handpiece always remove the bur tool from it Always make sure that you do not let the handpiece fall down or exposes it to any shocks The ball bearings and the motor could be damaged by this Only work with tools which are flawless Do not work with tools which are bent or out of balance whose shaft is worn out or are rusty Working with faulty tools will cause a stronge...

Page 13: ...e if it is bent Chuck is dirty Clean the chuck according to the instructions in the manual Bur does not clamp or does not spin Chuck is not tightened well Tighten the chuck firmly according to the instruction in the manual Knob of the handpiece can not be moved easily Chuck is dirty Clean the chuck according to the instructions in the manual Unit does not work at all Pull out the mains plug to res...

Page 14: ...n relates are conform to the following standards or normative documents 2014 30 EU directive EMC 2014 35 EU low voltage directive hadewe GmbH Grambartstraße 10 30165 Hannover Germany Symbols Caution Important Note Follow instructions for use I O Unit ON OFF Speed Regulation Up Down Switch to reverse direction LED ON LED is on in reverse direction Direct current Safety fuse Alternating Current Do n...

Page 15: ......

Page 16: ...hadewe GmbH Grambartstraße 10 30165 Hannover Germany info hadewe de www hadewe de ...

Reviews: