Safety Instructions
Rinse the tank daily and fill it with fresh water every day.
Empty the unit if it is not in use for a longer period of time
(e.g. before vacations). Clean out the tank. Set the spray
function on maximum. Now start the spray function, and the water
will spray for approx. 3 min. The fluid will be blown out of the
complete spray system and handpiece.
Ensure that the pa
tient’s foot and the handpiece are held
firmly in place during the operation. To prevent injury, work
with extreme caution on the foot of the patient, and use only light
pressure
To ensure that the burr/ tool sits firmly in the chuck, do not
work with tools which:
have an oily shaft
are worn
are bent.
Before using the equipment, assure that:
the tool can be pushed far enough into the chuck
(Test this with a dipstick and clean if necessary)
the inserted tool sits properly in the chuck and will not
cant/jam while clamping it
the tool is approved for using it at the max speed (see
manufacturer´s instructions), or only use the speed re-
quired by the instrument
the tool is absolutely dry.
For safety reasons do not change the tool while the motor
is running.