25
Techniqual Details
Safety Check
A safety check must be done in
compliance with the regulations of
the country where the device is
used. The inspection interval is
once a year.
Repeated control and control after
repair (DIN EN 62535, VDE 0751-1)
Inspection interval for repeated
control: 1 year
Technical Information:
Safety class: I
Application:
Type
B
(separate
measuring of discharge current not
necessary)
Power
connection:
NPS
(non-
detachable power connection)
The protective ground conductor is
not connected to touchable conduc-
tive parts.
Necessary
function
inspection:
speed,
display,
on/off,
suc-
tion/spray, clamping system.
Inspection before use
During the production control the
following inspections as per DIN EN
62353 (VDE 0751-1) are conducted
among others:
Optical control, unit discharge cur-
rent <1000µA, discharge current at
applied part <1000µA
Functional inspection of display,
on/off-button, suction/spray, speed
and clamping system. Safety and
function deficiencies have not been
found. By providing this information
an inspection before use can be
renounced as per DIN EN 62535,
which otherwise would have to be
ordered by the user.
Accessories
Art. No.
Description
4571
Protection washer
3497
Spray filter
5179
Micro filter
5116
filter
0970
Foot switch
8555
AQUAHY
The clamping system is designed
for rotating instruments, which have
a shaft of 2,35mm in accordance
with ISO EN ISO 179-1.
Summary of Contents for 0572
Page 1: ...Art No 0572 FB 04 278 Rev 0 Date 10 05 2016 Manual Argos 0572 ...
Page 14: ...14 Application ...
Page 26: ......
Page 27: ......
Page 28: ...hadewe GmbH Grambartstraße 10 30165 Hannover Germany info hadewe de www hadewe de ...