Before the use
Operation
The working area must conform to
the conditions described. Set up the
unit in such a way, that the air slits
are not blocked and the extracted
air can escape well. To guarantee
unhindered air flow there must a
gap of at least 12cm to the right of
the device. Position the device so
that the operating buttons are easy
to reach and the outlet of the hose
is a maximum distance of 60cm
from the patient's foot. The unit
should be 30 - 90cm from the floor.
Image: Example of positioning of foot care unit
in treatment room.
Image: Bird's eye view of work area
If you have a foot switch insert its
plug into the port on the right of the
device which displays this sym-
bol
.
Make sure, that the tank is filled
with spray liquid.
Check that the filter bag and dust
filter are correctly fitted.
Ensure that the tool is fixed.
On the back of the device you will
find the main power switch. Press
this switch to turn the unit on.
Take up the handpiece
To use the device press the button
(7) on the front of the unit.