38
ACP
probe
Steam
autoclave;
30 minutes under 115 degrees (C)
20 minutes under 121 degrees (C)
10 minutes under 134 degrees (C)
Do not expose the instrument to temperatures exceeding 134
degrees (C).
Sterilization should be performed in accordance with
instructions of the sterilization equipment.
Drying:
Sterilizable probes except ACP probe
No particular requirements.
ACP
probe
Dry it well after the sterilization.
Maintenance:
No particular requirements.
Inspection and Function Testing:
No cracks nor contaminations in appearance.
Connect the probe to main unit and make sure if you hear Doppler
sounds properly when you rub probe tip.
Packaging:
No particular requirements.
Storage:
No particular requirements.
Manufacturer contact:
Hadeco,
Inc.
2-7-11 Arima, Miyamae-ku, Kawasaki, 216-0003, Japan
Tel:
+81-44-877-4361
Fax:
+81-44-855-7301
The instructions provided above have been validated by the medical device
manufacturer as being CAPABLE of preparing a medical device for reuse. It remains the
responsibility of the processor to ensure that the processing as actually performed using
equipment, materials and personal in the processing facility achieve the desired result.
This requires validation and routine monitoring of process. Likewise any deviation by the
processor from the instruction provided should be evaluated for effectiveness and
potential adverse consequences.