24
h. UPPER (Upper limit for FHR)
If heart rate exceeds the upper limit for more than 30 sec, LCD will start flashing for
warning.
(1) Press Up and Down to select the upper limit in 5 BPM steps and press the button to
fix it.
i. LOWER (Lower limit for FHR)
If heart rate gets below lower limit for more than 30 sec, LCD will start flashing for
warning.
(1) Press Up and Down to select the lower limit in 5 BPM steps and press the button
to fix it.
j. PATIENT (Patient data input)
The patient data can be input. This setting contains ID, NAME, SEX, AGE, examination
DATE, anatomical SITE and MEMOrandum.
Save the patient data on Measurement mode and it will apply to all of future
measurement data until it's changed so that you won't have to input the same data for the
same patient again. If it is done on Freeze mode, the patient data will apply only to the
stored/FREEZE data selected.
(1) Prior to the patient data input, go to
OTHERS-DATE menu and set the date and time
for an initial setting.
Scroll to the item where you wish to input and
press Enter to display data input screen. With
SEX, pressing Enter changes Male to Female or
vice versa.
DATE data can be changed on PATIENT menu
when on FREEZE mode only.
Upper limit
Lower limit