Operation—4–20 mA/HART
37
If range is set to 000.0% select solution (Conc)
o1.ConC
-01- NaCl (default)
Solution (Conc)
(refer to
for more
information)
-02- HCI
-03- NaOH
-04- H
2
SO
4
-05- HNO
3
-06- H
2
SO
4
-07- HCI
-08- HNO
3
-09- H
2
SO
4
-10- NaOH
Select characteristic (Linear / Logarithmic curve)
o1.CHAR
2
LIN (default)
Linear/Logarithmic curve
LOG
Specify the value for the 4 and 20 mA signals
o1. 4mA
000.0 mS
LIN: Enter current start
o1.20mA 000.0 mS
LIN: Enter current end
o1. 4mA
1 mS (default)
LOG: Enter current start
o1.20mA 100 mS (default)
LOG: Enter current end
Set time averaging filter for reducing noise
o1.FtME
0 SEC (default)
Time constant of output filter (refer to
Select a 22 mA signal during errors
o1.FAIL
ON
OFF (default)
Signal behavior during HOLD
o1.HOLD
LAST (default)
During HOLD the last measured value is
maintained at the output
FIX
During HOLD a value (to be entered) is
maintained at the output
5.2.1
Current output configuration (continued)
Summary of Contents for si792 T
Page 2: ......
Page 12: ...Specifications 10...
Page 32: ...Installation 30...
Page 36: ...Interface and navigation 34...
Page 66: ...Operation Foundation Fieldbus 64...
Page 80: ...Maintenance 78...
Page 86: ...Troubleshooting 84...
Page 88: ...Replacement parts 86...
Page 94: ...Warranty liability and complaints 92...
Page 98: ...Calibration solutions 96...
Page 106: ...Concentration curves 104...
Page 112: ...ATEX safety instructions 110 Figure 20 Sanitary style sensor...
Page 116: ...Index 114...