Operation—Foundation Fieldbus
63
6.2.4 Calibration via Foundation Fieldbus
The transmitter can be calibrated via Foundation Fieldbus using the
comparison or grab sample method.
1.
Make sure the system is configured for conductivity
(PRIMARY_VALUE_TYPE = mS/cm or S/m).
2.
Collect a grab sample and set CAL_SAMPLE_PRD to sample. The
conductivity value of the sample is stored. After writing, the
parameter is automatically reset to NOP (no operation).
3.
Read the parameter CAL_SAMPLE_PRD_STORED_VAL. It
contains the stored value.
4.
Measure the grab sample and write the lab value in
CAL_PRODUCT. The device is now calibrated.
CAL_SAMPLE_PRD_STORED_VAL is reset to zero.