Operation—Foundation Fieldbus
48
GRANT_DENY
Grant 0
R/W
Deny 0
R/W
HARD_TYPES Hardware
type
1
R
RESTART Restart
R/W
FEATURES
Feature supported
Reports/ Soft W Lock
R
FEATURES
Feature selected
Reports/ Soft W Lock
R/W
CYCLE_TYPE Cycle
type
Scheduled/
Block Execution
R
CYCLES_SEL Cycle
selected
Scheduled/
Block Execution
R/W
MIN_CYCLE_T Min
cycle
time
1600
1
/
32
ms (50 ms)
R
MEMORY_SIZE Memory
size
R
NV_CYCLE_T
Non-volatile cycle time
R
FREE_SPACE Free
space
R
FREE_TIME Free
time
R
SHED_RCAS R/W
SHED_ROUT R/W
FAULT_STATE Fault
state
R
SET_FSTATE Set
fault
state
1
R/W
CLR_FSTATE Clear
fault
state
1
R/W
MAX_NOTIFY Max
notifications 20
R
LIM_NOTIFY Limit
of
notification 8
R/W
CONFIRM_TIME Confirmation
time
640000
1
/
32
ms
R/W
WRITE_LOCK Write
locking
1
(unlocked)
R/W
UPDATE_EVT
Unacknowledged 0
R/W
Update state
0
R
Time stamp
0
R
Static revision
0
R
Relative index
0
R/W
Table 7 Bus parameters—resource block (RB) (continued)
Parameter
Description
Default
R/W
Summary of Contents for si792 T
Page 2: ......
Page 12: ...Specifications 10...
Page 32: ...Installation 30...
Page 36: ...Interface and navigation 34...
Page 66: ...Operation Foundation Fieldbus 64...
Page 80: ...Maintenance 78...
Page 86: ...Troubleshooting 84...
Page 88: ...Replacement parts 86...
Page 94: ...Warranty liability and complaints 92...
Page 98: ...Calibration solutions 96...
Page 106: ...Concentration curves 104...
Page 112: ...ATEX safety instructions 110 Figure 20 Sanitary style sensor...
Page 116: ...Index 114...