![Hach sensIONTM+ pH31 User Manual Download Page 9](http://html1.mh-extra.com/html/hach/sensiontm-ph31/sensiontm-ph31_user-manual_3252734009.webp)
Option
Description
Cal. frequency
Calibration reminder—can be set between 0–7 days
(default daily). The display shows the remaining time to
the new calibration. Refer to
Set the calibration reminder
on page 10 for more information.
Display mV
Display mV—select YES or NO to show mV.
Standard 220 mV
Standard 220 mV—calibrate with a 220 mV standard
solution
To a X value
To a X value—use a specific standard solution. Adjust
the value during the calibration.
Data introduction
Data introduction Compensation—enter the
compensation value
Factory adjust
Factory adjust—enter the compensation value
Calibration types
Different calibration types can be selected.
1.
From the main menu use the or to select CALIBRATION.
Confirm.
2.
Use the to enter the calibration menu.
3.
Use the or to select Calibration type.
Option
Description
Technical buffers
pH 2.00, 4.01, 7.00, 9.21 and 10.01 at 25 °C
(77 °F)
DIN19266 Buffers
pH 1.679, 4.006, 6.865, 7.000, 9.180, 10.012 and
12.454
User Buffers
Selected when the technical or DIN19266 buffers
are not used. Refer to
Technical buffer solutions
(DIN 19267)
on page 18 for pH values of
specific buffer sets at varying temperatures.
Calibration to a X value
To adjust manually any scale value of the
measured pH.
Option
Description
Data introduction
Manual probe constant introduction.
Theoretical calibration
The probe calibration data is replaced at 25 °C
(77 °F).
Calibration procedure
This procedure is for general use with liquid calibration solutions. Refer
to the documents that are included with each probe for additional
information.
Note: Solutions should be stirred during calibration. For more information about the
stirring settings, refer to
Change the stirring settings
on page 12.
1.
Pour the buffers or calibration solutions into the labeled calibration
beakers.
2.
From the main menu use the or and and to select the
CALIBRATION parameter. Confirm.
3.
If required select the Operator ID (1 to 10) and confirm.
4.
Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the
membrane.
5.
Push to Start calibration.
6.
Push to measure the first calibration solution.
The next calibration solution is shown.
7.
Rinse the probe with deionized water and put the probe into the
second calibration beaker. Be sure that there are no air bubbles in
the membrane.
8.
Push to measure the second calibration solution.
The next calibration solution is shown.
9.
Rinse the probe with deionized water and put the probe into the third
calibration beaker. Be sure that there are no air bubbles in the
membrane.
10.
Push to measure the third calibration solution.
When the calibration is good, the display briefly shows Calibration
OK and then returns to the main menu.
Note: When a printer is connected the print menu opens and the result can be
printed.
English
9
Summary of Contents for sensIONTM+ pH31
Page 2: ...English 3 Fran ais 19 Espa ol 37 Portugu s 54 72 86 102 117 2...
Page 5: ...1 2 3 4 English 5...
Page 15: ...1 2 English 15...
Page 22: ...1 2 3 4 22 Fran ais...
Page 33: ...1 2 Fran ais 33...
Page 50: ...1 2 50 Espa ol...
Page 57: ...1 2 3 4 Portugu s 57...
Page 68: ...1 2 68 Portugu s...
Page 74: ...1 2 3 4 74...
Page 75: ...GFCI GFI 1 2 2 3 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 3 75...
Page 82: ...1 2 82...
Page 88: ...1 2 3 4 88...
Page 89: ...AC GFCI GFI AC 1 2 2 3 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 89...
Page 90: ...3 AC 1 5 2 6 3 7 4 ID ID 90...
Page 97: ...99 pH 1 1 pH 0 1 N HCl 97...
Page 98: ...1 2 98...
Page 104: ...1 2 3 4 104...
Page 105: ...AC GFCI GFI AC 1 2 2 3 AC 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 3 AC 105...
Page 112: ...114 pH 1 1 pH 0 1 N HCl 112...
Page 113: ...1 2 113...
Page 114: ...2 58 mV 58mV 112 a 70 t 100 112 2 112 112 3 pH 12 78 19 C pH 112 C 60 112 150 300 114...
Page 119: ...1 2 119...
Page 120: ...3 4 AC Ground Fault Circuit Interrupt GFCI GFI AC AC 1 2 2 3 AC 3 4 120...
Page 121: ...2 1 5 DIN 2 6 3 RS 232 7 pH 4 8 3 AC 121...
Page 127: ...3 17 40 GLP 17 40 4 LabCom Easy 125 17 129 1 pH 1 pH 0 1 N HCl 127...
Page 128: ...1 2 128...
Page 132: ......
Page 133: ......