Troubleshooting
Refer to the following table for common problem messages or
symptoms, possible causes and corrective actions.
Table 2 Calibration warnings and errors
Error/Warning
Solution
Asymmetry >
58 mV
Slope out of range (accepted values ± 58 mV).
Calibrate again.
Connect a new probe.
Buffer or electrode
in poor conditions.
Calibrate again.
Examine the probe: Clean the probe (refer to
Clean the
probe
on page 13 for more information); make sure that
there are no air bubbles in the membrane. Shake the probe
like a thermometer; Connect a different probe to verify if
problem is with probe or meter.
Examine the buffer solution: Make sure that the buffer used
matches the buffer specified in configuration; make sure of
the temperature specification in configuration; use new
buffer solution.
Sens. (a) < 70%
BUFFER NOT
RECOGNIZED
UNSTABLE
READING
Time t> 100 s
Calibrate again.
Examine the probe: Clean the probe (refer to
Clean the
probe
on page 13 for more information); make sure that
there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Table 2 Calibration warnings and errors (continued)
Error/Warning
Solution
Electrode in poor
conditions.
Examine the probe: Clean the probe (refer to
Clean the
probe
on page 13 for more information); make sure that
there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Check the
electrode
SAME BUFFERS
Calibrate again.
Examine the probe: Clean the probe (refer to
Clean the
probe
on page 13 for more information); make sure that
there are no air bubbles in the membrane. Shake the probe
like a thermometer; connect a different probe to verify if
problem is with probe or meter.
Examine the buffer solution: Use new buffer solution.
Table 3 Measurement warnings and errors
Error/Warning
Solution
pH 12.78 19°C
Calibrate again
pH out of range
Examine the probe: Clean the probe (refer to
Clean the
probe
on page 13 for more information); make sure that
there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.
Temp out of range
°C
Examine the temperature sensor.
Connect a different probe to verify if problem is with probe
or meter.
Time > 60 s
Make sure that the membrane and the diaphragm are
properly immersed in the sample.
Examine the temperature.
Examine the probe: Clean the probe (refer to
Clean the
probe
on page 13 for more information); make sure that
there are no air bubbles in the membrane. Shake the
probe like a thermometer; connect a different probe to
verify if problem is with probe or meter.
Time > 150 s
Time > 300 s
16
English
Summary of Contents for sensIONTM+ pH31
Page 2: ...English 3 Fran ais 19 Espa ol 37 Portugu s 54 72 86 102 117 2...
Page 5: ...1 2 3 4 English 5...
Page 15: ...1 2 English 15...
Page 22: ...1 2 3 4 22 Fran ais...
Page 33: ...1 2 Fran ais 33...
Page 50: ...1 2 50 Espa ol...
Page 57: ...1 2 3 4 Portugu s 57...
Page 68: ...1 2 68 Portugu s...
Page 74: ...1 2 3 4 74...
Page 75: ...GFCI GFI 1 2 2 3 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 3 75...
Page 82: ...1 2 82...
Page 88: ...1 2 3 4 88...
Page 89: ...AC GFCI GFI AC 1 2 2 3 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 89...
Page 90: ...3 AC 1 5 2 6 3 7 4 ID ID 90...
Page 97: ...99 pH 1 1 pH 0 1 N HCl 97...
Page 98: ...1 2 98...
Page 104: ...1 2 3 4 104...
Page 105: ...AC GFCI GFI AC 1 2 2 3 AC 3 4 2 1 5 PC DIN 2 6 3 PC RS 232 7 pH 4 8 3 AC 105...
Page 112: ...114 pH 1 1 pH 0 1 N HCl 112...
Page 113: ...1 2 113...
Page 114: ...2 58 mV 58mV 112 a 70 t 100 112 2 112 112 3 pH 12 78 19 C pH 112 C 60 112 150 300 114...
Page 119: ...1 2 119...
Page 120: ...3 4 AC Ground Fault Circuit Interrupt GFCI GFI AC AC 1 2 2 3 AC 3 4 120...
Page 121: ...2 1 5 DIN 2 6 3 RS 232 7 pH 4 8 3 AC 121...
Page 127: ...3 17 40 GLP 17 40 4 LabCom Easy 125 17 129 1 pH 1 pH 0 1 N HCl 127...
Page 128: ...1 2 128...
Page 132: ......
Page 133: ......