59
SECTION 6
DISSOLVED OXYGEN IN WATER
(0 to 20 mg/L
)
1.
Assemble the
dissolved oxygen probe
as described in
Section
2.
At least one hour
before measurement,
polarize the probe by
connecting it to the
meter. See
Section 3.3.2
.
3.
Zero the
sens
ion
™
6
Dissolved
Oxygen meter prior to
calibration when
measuring dissolved
oxygen levels less than
1 mg/L or 10%
saturation.
4.
Secure the probe
cable to the calibration
and storage chamber by
wrapping cable through
the bottom of the
chamber lid before
filling with water.