43
SECTION 6
TROUBLESHOOTING
6.1
Introduction
Correcting problem conditions with the sens
ion
™
electrochemical meters is limited to responding to the error
message displayed. Other problems must be handled by a Hach
technician at a service center. Refer to Repair Service. Do not
attempt to service the meter as there are no field-serviceable
parts. Opening the meter case will void the warranty.
6.2
Shorting Test
This test detects the meter offset
1. Turn the meter on. Connect the shorting cap to the BNC
connector (see Figure 6). Select the BNC connector from
Setup 1
.
2. Change the output to display in mV (press the
pH/mV
key to
toggle between mV and pH). The display should show
0.1
mV
. If it does not, contact Hach Service.
Figure 6
Attaching Shorting Caps to the BNC Connector