11
Tref:
20
˚
C
25
˚
C
Other temperature
Measuring conditions
Measurement:
By stability
In continuous
By time
Measuring conditions
Measurement:
By stability
In continuous
By time
Measuring conditions
Measurement:
By stability
In continuous
By time
Measuring conditions
TC: lineal 2.00%/
˚
C
Reference T: 25
˚
C
Meas.: stability
Start measusrement
Measuring conditions
TC:
Lineal
Natural waters
Measuring conditions
TC lineal:
2
.00%/
˚
C
Measuring conditions
TC: lineal 2.00%/
˚
C
Reference T: 25
˚
C
Meas.: stability
Start measusrement
Measuring conditions
TC: lineal 2.00%/
˚
C
Reference T: 25
˚
C
Meas.: stability
Start measusrement
Measuring conditions
TC: lineal 2.00%/
˚
C
Reference T: 25
˚
C
Meas.: stability
Start measusrement
Measuring conditions
Measuring conditions
In continuous
Acquis. Interval
01
:00
Measuring conditions
By time
Acquis. Interval
00:
01
:00
Measuring conditions
In continuous
Acquis. Interval
01:
30
Measuring conditions
By time
Measurement time
00:
01
:00
...
Start meas. EC
By stability
10-04-2010 17:45
EC MEASURE
New calibration
in 15 days
Default programming (standard)
Other options
TC
•
linear, TC = 2.00 % /
ºC
.
• linear, TC selectable between
0.00 to 9.99
%/ºC.
• non-linear for natural waters, according to EN 27888
TRef
•
25 ˚C
• 20 ˚C
• 25 ˚C
• other values
Measuring mode
•
By stability
.
-
Criterion, 1 digit in 6 s.
•
In continuous
.
- The result is stored when the measurement
has finished.
- The result is printed when the measurement
has finished.
•
By stability.
-
Criterion, 1 digit in 6 s.
•
In continuous.
Selection of the time interval for:
-
Data
storage.
-
Data
printing.
•
By time.
Selection of the time interval for:
-
Data
storage.
-
Data
printing.
“Re-programming” flow chart
ENGLISH
Important note:
Once the measurement has been re-programmed, it will be
performed in accordance with the new parameters.
4. Operation
4.8. Re-programming EC measurements
Start of measurement
with new programming
Data storage in
the Data Logger. It
coincides with the
printing and PC
sending. Select-
ing 0 store/print
at the end of the
measurement.