18
SYSTEM
Sample ID
Data output
Type of report
SYSTEM
Language
Constrast adjust
Readjust temp.
SYSTEM
Readjust temperature
Correction
25.0
˚
C: 0.1
˚
C
85.0
˚
C: 0.0
˚
C
Save changes
SYSTEM
Readjust temperature
Correction
25.0
˚
C: 0.1
˚
C
85.0
˚
C: 0.0
˚
C
Save changes
SYSTEM
Readjust temperature
Correction
25.0°C: 0.0 °C
85.0°C: 0.0 °C
SYSTEM
Readjust temperature
Correction at 25.0
˚
C
0
.0 °C
SYSTEM
10-04-2007 17:15
°C °F
µS/cm
µS/cm
mS/cm
mS/cm
15.0 59
119 1147 10.48
92.5
16.0 60.8
122
1173 10.72
94.4
17.0 62.6
125
1199 10.95
96.3
18.0 64.4
127
1225 11.19
98.2
19.0 66.2
130
1251 11.43
100.1
20.0
68
133 1278
11.67
102.1
21.0
69.8
136
1305
11.91
104.0
22.0
71.6
139
1332
12.15
105.4
23.0
73.4
142
1359
12.39
107.9
24.0
75.2
145
1386
12.64
109.8
25.0
77
147
1413
12.88
111.8
26.0
78.8
150
1440
13.13
113.8
27.0
80.6
153
1467
13.37
115.7
28.0
82.4
156
1494
13.62
--
29.0
84.2
159
1522
13.87
--
30.0
86
162
1549
14.12
--
31.0
87.8
165
1581
14.37
--
32.0
89.6
168
1609
14.62
--
33.0
91.4
171
1638
14.88
--
34.0
93.2
174
1667
15.13
--
35.0
95
177
1696
15.39
--
4. Operation
4.13. Temperature readjustment
The
sensION
™
+
EC 71
allows the correction of the deviation
presented by a temperature probe (integrated in the electrode or
separated sensor) at 25º and 85ºC. In this way the EC-meter can
be used as a precise thermometer.
4.14. Recognized standards
4.14.1. Molar Standards
Table of conductivity values according to temperature
Values stored in the memory of
sensION
™
+
EC 71
.
EC71 HACH LANGE 7010.indd ANG:18
EC71 HACH LANGE 7010.indd ANG:18
8/7/10 16:14:57
8/7/10 16:14:57