background image

Specifications

Specifications are subject to change without notice.

Specification

Details

Dimensions (W x D x H)

37 x 30 x 170 mm (1.45 x 1.18 x 6.69 in.)

Enclosure rating

IP67 

Weight

135 g (0.297 lb) with batteries

Power requirements (internal)

AAA alkaline batteries (4x)

Battery life

450 hours

Operating temperature

0 to 50 °C (32 to 122 °F)

Storage temperature

–20 to 60 °C (–4 to 140 °F) )

Operating humidity

80% (non-condensing)

Altitude

2000 m (6562 ft)

Instrument range

Cond

LR

: 0 to 1999 µS/cm; Cond

HR

: 0 to 19.99 mS/cm

Resolution

Cond

LR

: 1 µS/cm; Cond

HR

: 0.01 mS/cm

Accuracy

Cond

LR

: ±1% FS

1

; Cond

HR

: ±2% FS

Note: Temperature calibration is not available.

Calibration points

Cond

LR

: 2 points (auto), 1 point (custom); Cond

HR

: 1 point (auto and custom)

Auto-recognition calibration standards

Cond

LR

: 147 µS/cm and 1413 µS/cm; Cond

HR

: 12.88 mS/cm

Certifications

CE mark, FCC, Industry Canada, KC Mark, RCM, China RoHS

Warranty

1 year for manufacturing faults only. Damage from use is not covered.

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or
consequential damages resulting from any defect or omission in this manual. The
manufacturer reserves the right to make changes in this manual and the products it
describes at any time, without notice or obligation. Revised editions are found on the
manufacturer’s website.

Safety information

N O T I C E

 

The manufacturer is not responsible for any damages due to misapplication or misuse of this
product including, without limitation, direct, incidental and consequential damages, and disclaims
such damages to the full extent permitted under applicable law. The user is solely responsible to
identify critical application risks and install appropriate mechanisms to protect processes during a
possible equipment malfunction.

Please read this entire manual before unpacking, setting up or operating this equipment.
Pay attention to all danger and caution statements. Failure to do so could result in serious
injury to the operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or
install this equipment in any manner other than that specified in this manual.

1

Based on a 2-point calibration and the conductivity range of the calibration standards includes the conductivity
value of the water sample.

2

Summary of Contents for Pocket Pro CondLR

Page 1: ...User Manual DOC022 52 80396 Pocket Pro CondLR CondHR 1...

Page 2: ...equential damages resulting from any defect or omission in this manual The manufacturer reserves the right to make changes in this manual and the products it describes at any time without notice or ob...

Page 3: ...plier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items for proper disposal Certification Canadian Radio Interference Causing Equ...

Page 4: ...tion the receiving antenna for the device receiving the interference 3 Try combinations of the above Product overview The Pocket Pro conductivityLR tester and conductivityHR tester measure the conduct...

Page 5: ...ical type and are inserted in the correct orientation Do not mix new and used batteries WA RNING Fire hazard Battery substitution is not permitted Use only alkaline batteries Install the four AAA alka...

Page 6: ...y power is less than 10 Shows when the lock feature is on When the lock feature is on the parameter value on the display does not change Push to set the lock feature to on or off Shows when the tester...

Page 7: ...standard CondLR tester One or two auto recognition calibration standards 1 Set the power to on 2 Remove the cap from the sensor 3 Push to go to calibration mode The auto recognition standard s the te...

Page 8: ...perature compensation factor linear 0 00 to 9 99 per C default 2 00 per C The correction factor may need to be identified experimentally For example the factor for ultrapure water is 4 55 per C and th...

Page 9: ...rature range of the tester Contact technical support as necessary Note The tester can still be used if a temperature sensor failure has occurred but without automatic temperature compensation Battery...

Page 10: ...ach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Com...

Reviews: