1—42
Chlorine, Free and Total, HR,
continued
4.
Press:
ZERO/SCROLL
The display will show
“- - - -” followed by “0.0”.
Remove the blank.
5.
Fill another 1-cm/10-mL
sample cell to the 5-mL line
with sample. Cap.
Note:
Do not use the same
sample cells for free and
total chlorine without
thoroughly rinsing the cells
between the free and total
tests.
6.
Add the contents of one
DPD
Free
Chlorine or one
DPD
Total
Chlorine Powder
Pillows to the sample cell
(the prepared sample). Cap
the cell and shake gently for
20 seconds.
Note:
Wipe liquid off sample
cells.