2—25
User-Entered Calibration,
continued
5.
Press
READ/ENTER
to delete. Repeat for all points to be deleted.
Note: The minimum number of valid points is two. For example, if five points have been
entered, three can be deleted using this feature.
6.
Press
MENU
to return to the measurement mode.
Retrieving the Factory Calibration
1.
Select the range to restore factory default calibration. See
Switching Ranges
on page 2—4
.
2.
Press and hold the
MENU
key until “USER”, then “CAL” appears.
3.
Press the
ZERO/SCROLL
key to find dFL.
4.
Press the
READ/ENTER
key to select dFL and restore the instrument to the
factory default calibration.
Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To
turn SCA back on, restore the meter to factory default calibration.