Specifications

Details

Minimum sample volume

25 mL

Minimum immersion depth

25.4 mm (1 in.)

Body material (standard)

Epoxy

Sensor material

Solid-state crystal

Reference electrolyte

Non-refillable Dritek gel reference element

Cable connection

M12 digital output and connector

Dimensions

Diameter: 12 mm (0.47 in.)
Length: 175 mm (6.9 in.) total; 103 mm (4.1 in.) below head
Cable length: ISECL18101: 1 m (3.3 ft); ISECL18103: 3 m (9.8 ft)

Warranty

1 year on the probe. This warranty covers manufacturing defects,
but not improper use or wear.

Certifications

CE, FCC/ISED

Section 3 Safety information

3.1 Intended use

The Intellical probes are intended for use by individuals who measure water quality parameters in the
laboratory. The Intellical probes do not treat or alter water.

3.2 Use of hazard information

D A N G E R

 

Indicates a potentially or imminently hazardous situation which, if not avoided, will result in death or
serious injury.

W A R N I N G

 

Indicates a potentially or imminently hazardous situation which, if not avoided, could result in death
or serious injury.

C A U T I O N

 

Indicates a potentially hazardous situation that may result in minor or moderate injury.

N O T I C E

 

Indicates a situation which, if not avoided, may cause damage to the instrument. Information that
requires special emphasis.

3.3 Precautionary labels

Read all labels and tags attached to the instrument. Personal injury or damage to the instrument
could occur if not observed. A symbol on the instrument is referenced in the manual with a
precautionary statement.

Electrical equipment marked with this symbol may not be disposed of in European
domestic or public disposal systems. Return old or end-of-life equipment to the
manufacturer for disposal at no charge to the user.

4

   

English

Summary of Contents for ISECL181

Page 1: ...DOC022 53 80030 ISECL181 05 2021 Edition 5 User Manual...

Page 2: ......

Page 3: ...on 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 6 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 6 3 Low level measurements 7 6 4 Interferences...

Page 4: ...Table of Contents 2...

Page 5: ...ications Details Probe type Digital combination probe with a non refillable reference junction and a built in temperature sensor Measurement range 0 1 mg L 3x10 6 M to 35 500 mg L 1 M Cl Accuracy 0 02...

Page 6: ...information D A N G E R Indicates a potentially or imminently hazardous situation which if not avoided will result in death or serious injury W A R N I N G Indicates a potentially or imminently hazar...

Page 7: ...e and time settings when the meter starts for the first time or after battery replacement 5 Connect the probe to the meter Section 5 Calibration The procedure that follows is applicable to meters that...

Page 8: ...eionized water Dry the probe with a lint free cloth 6 Put the probe in the standard solution with the sensor fully submerged Do not put the probe on the bottom or sides of the beaker 7 Shake the probe...

Page 9: ...when the meter is set to read continuously Air bubbles below the sensor can cause a slow response or error in the measurement Make sure to remove air bubbles before and during measurements 6 2 Sample...

Page 10: ...on relative to the primary ion High concentrations of ions that form insoluble salts with silver will stay on the sensing element and cause measurement errors Strong reducing substances will form a su...

Page 11: ...e bottom or sides of the beaker 7 Shake the probe from side to side to remove air bubbles 8 Stir for 30 to 60 seconds then read the value of the standard solution The meter accepts or rejects the resu...

Page 12: ...junction with deionized water Shake the probe down to remove air bubbles The probe is not conditioned to the sample sufficiently Condition the probe Refer to Preparation for use on page 5 Sample prope...

Page 13: ...f standard solution to the sample Refer to Table 2 Table 2 Standard solution volumes and concentrations Measured sample concentration Volume of standard to add Concentration of standard solution 1 to...

Page 14: ...2370853 Chloride Standard Solution 1000 mg L as Cl 500 mL 18349 Wash bottle polyethylene 500 mL 1 62011 Disposable wipes 11 x 22 cm 280 pkg 2097000 Beaker 30 mL plastic colorless 80 pkg SM5010 Beaker...

Page 15: ......

Page 16: ...932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 rout...

Reviews: