8.2 Storage
Keep the sensor protection cap on the probe during storage to prevent damage to the sensing
element.
Note: Make sure to condition the probe after long-term storage. Refer to
on page 5.
1.
Rinse the probe with deionized water. Dry the probe with a lint-free cloth.
2.
Make sure that the probe sensor is dry, then install the sensor protection cap. Refer to
on page 3.
Section 9 Troubleshooting
Keep the sensor protection cap on the probe when not in use for the best accuracy, stabilization time
and life of the probe. To check the probe performance, refer to
on page 11. To check
the accuracy of sample measurements, refer to
on page 11.
Problem
Possible cause
Solution
Decreased probe performance
causes slow stabilization and
prevents accurate calibrations or
measurements.
The probe has contamination
on the sensing element.
Clean and condition the probe.
Refer to
on page 9.
The reference junction is
clogged.
Fully rinse the reference junction
with deionized water. Shake the
probe down to remove air
bubbles.
The probe is not conditioned
to the sample sufficiently.
Condition the probe. Refer to
Sample properties cause slow
stabilization or inaccurate
measurements.
The sample temperature is
low, or there is a large
temperature difference
between samples.
Increase the sample
temperature or adjust the
temperature of different samples
to be the same (within 2 °C
(3.6 °F)).
Procedure problem causes slow
stabilization and prevents
accurate calibrations or
measurements.
Air bubbles are around or
below the probe tip.
Carefully tap or shake the probe
to remove air bubbles.
The ISA was not added.
Add one ISA powder pillow to
each 25 mL of sample and
standard solution.
The stir speed is too slow or
too fast.
Try a different stir speed.
Magnetic stirrers can
become warm and increase
the solution temperature.
Put a piece of insulating material
between the stirrer and beaker.
An incorrect standard
solution was used or the
standard solution has
contamination.
Use the specified standard
solution of good quality.
Too much light is on the
sensing element.
Put a cover on the beaker to
give protection from direct light
during measurement.
10
English