8.2 Storage

Keep the sensor protection cap on the probe during storage to prevent damage to the sensing
element.

Note: Make sure to condition the probe after long-term storage. Refer to 

Preparation for use

 on page 5.

1.

Rinse the probe with deionized water. Dry the probe with a lint-free cloth.

2.

Make sure that the probe sensor is dry, then install the sensor protection cap. Refer to 

Product

overview

 on page 3.

Section 9 Troubleshooting

Keep the sensor protection cap on the probe when not in use for the best accuracy, stabilization time
and life of the probe. To check the probe performance, refer to 

Slope check

 on page 11. To check

the accuracy of sample measurements, refer to 

Standard additions check

 on page 11.

Problem

Possible cause

Solution

Decreased probe performance
causes slow stabilization and
prevents accurate calibrations or
measurements.

The probe has contamination
on the sensing element.

Clean and condition the probe.
Refer to 

Clean the probe

on page 9.

The reference junction is
clogged.

Fully rinse the reference junction
with deionized water. Shake the
probe down to remove air
bubbles.

The probe is not conditioned
to the sample sufficiently.

Condition the probe. Refer to 

Preparation for use

 on page 5.

Sample properties cause slow
stabilization or inaccurate
measurements.

The sample temperature is
low, or there is a large
temperature difference
between samples.

Increase the sample
temperature or adjust the
temperature of different samples
to be the same (within 2 °C
(3.6 °F)).

Procedure problem causes slow
stabilization and prevents
accurate calibrations or
measurements.

Air bubbles are around or
below the probe tip.

Carefully tap or shake the probe
to remove air bubbles.

The ISA was not added.

Add one ISA powder pillow to
each 25 mL of sample and
standard solution.

The stir speed is too slow or
too fast.

Try a different stir speed.

Magnetic stirrers can
become warm and increase
the solution temperature.

Put a piece of insulating material
between the stirrer and beaker.

An incorrect standard
solution was used or the
standard solution has
contamination.

Use the specified standard
solution of good quality.

Too much light is on the
sensing element.

Put a cover on the beaker to
give protection from direct light
during measurement.

10

   

English

Summary of Contents for ISECL181

Page 1: ...DOC022 53 80030 ISECL181 05 2021 Edition 5 User Manual...

Page 2: ......

Page 3: ...on 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 6 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 6 3 Low level measurements 7 6 4 Interferences...

Page 4: ...Table of Contents 2...

Page 5: ...ications Details Probe type Digital combination probe with a non refillable reference junction and a built in temperature sensor Measurement range 0 1 mg L 3x10 6 M to 35 500 mg L 1 M Cl Accuracy 0 02...

Page 6: ...information D A N G E R Indicates a potentially or imminently hazardous situation which if not avoided will result in death or serious injury W A R N I N G Indicates a potentially or imminently hazar...

Page 7: ...e and time settings when the meter starts for the first time or after battery replacement 5 Connect the probe to the meter Section 5 Calibration The procedure that follows is applicable to meters that...

Page 8: ...eionized water Dry the probe with a lint free cloth 6 Put the probe in the standard solution with the sensor fully submerged Do not put the probe on the bottom or sides of the beaker 7 Shake the probe...

Page 9: ...when the meter is set to read continuously Air bubbles below the sensor can cause a slow response or error in the measurement Make sure to remove air bubbles before and during measurements 6 2 Sample...

Page 10: ...on relative to the primary ion High concentrations of ions that form insoluble salts with silver will stay on the sensing element and cause measurement errors Strong reducing substances will form a su...

Page 11: ...e bottom or sides of the beaker 7 Shake the probe from side to side to remove air bubbles 8 Stir for 30 to 60 seconds then read the value of the standard solution The meter accepts or rejects the resu...

Page 12: ...junction with deionized water Shake the probe down to remove air bubbles The probe is not conditioned to the sample sufficiently Condition the probe Refer to Preparation for use on page 5 Sample prope...

Page 13: ...f standard solution to the sample Refer to Table 2 Table 2 Standard solution volumes and concentrations Measured sample concentration Volume of standard to add Concentration of standard solution 1 to...

Page 14: ...2370853 Chloride Standard Solution 1000 mg L as Cl 500 mL 18349 Wash bottle polyethylene 500 mL 1 62011 Disposable wipes 11 x 22 cm 280 pkg 2097000 Beaker 30 mL plastic colorless 80 pkg SM5010 Beaker...

Page 15: ......

Page 16: ...932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 rout...

Reviews: