Hach ISECL181 User Manual Download Page 13

9.1 Slope check

Use the mV value of two standard solutions to make sure the probe gives the correct slope.

1.

Prepare two standard solutions that are ten times apart in concentration (e.g., 100 mg/L and
1000 mg/L Cl

). Select standard solutions with a concentration above and below the typical

sample concentration. Use a minimum concentration of 7 mg/L.

2.

Use the measurement procedure to add the ISA and measure the mV value of each standard
solution.

3.

Calculate the difference in the mV value of the two standard solutions to find the slope. If the
probe is in good condition, the slope will be 58 mV (within the ± slope limits of the method) at
25 °C (77 °F).

9.2 Standard additions check

To make sure that the sample measurement is accurate, add a small volume of a standard solution
to the sample and calculate the percent recovery. The sample with the known volume of standard
solution is known as a spiked sample.

1.

Use the measurement procedure to measure the concentration of a 25-mL sample.

2.

Use a pipet to add the applicable volume of standard solution to the sample. Refer to 

Table 2

.

Table 2  Standard solution volumes and concentrations

Measured sample

concentration

Volume of standard to

add

Concentration of standard

solution

1 to 2 mg/L

0.5 mL

100 mg/L Cl

3 to 6 mg/L

1.0 mL

100 mg/L Cl

7 to 15 mg/L

0.3 mL

1000 mg/L Cl

15 to 30 mg/L

0.5 mL

1000 mg/L Cl

30 to 60 mg/L

1.0 mL

1000 mg/L Cl

3.

Measure the concentration of the spiked sample.

4.

Calculate the expected (theoretical) concentration of the spiked sample:
C

E

 = (C

S

 x V

S

/V

T

) + (C

SS

 x V

SS

/V

T

)

Where:

• C

E

 = expected (theoretical) concentration of the spiked sample

• C

S

 = concentration of the sample (mg/L) before the standard solution was added

• C

SS

 = concentration of the standard solution (mg/L)

• V

S

 = sample volume (mL) before the standard solution was added

• V

SS

 = volume of the standard solution (mL)

• V

T

 = total volume (standard solution volume (mL) + sample volume)

5.

Calculate the percent recovery of the standard addition. A percent recovery of 100 (±5)% is an
indication that the sample measurements are accurate.
Percent recovery = C

M

/C

E

 x 100 

Where:

• C

M

 = measured concentration of the sample after the addition of the standard solution

• C

E

 = expected (theoretical) concentration of the sample after the addition of the standard

solution

English

   

11

Summary of Contents for ISECL181

Page 1: ...DOC022 53 80030 ISECL181 05 2021 Edition 5 User Manual...

Page 2: ......

Page 3: ...on 5 5 1 Calibration notes 5 5 2 Calibration procedure 6 Section 6 Sample measurement 6 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 6 3 Low level measurements 7 6 4 Interferences...

Page 4: ...Table of Contents 2...

Page 5: ...ications Details Probe type Digital combination probe with a non refillable reference junction and a built in temperature sensor Measurement range 0 1 mg L 3x10 6 M to 35 500 mg L 1 M Cl Accuracy 0 02...

Page 6: ...information D A N G E R Indicates a potentially or imminently hazardous situation which if not avoided will result in death or serious injury W A R N I N G Indicates a potentially or imminently hazar...

Page 7: ...e and time settings when the meter starts for the first time or after battery replacement 5 Connect the probe to the meter Section 5 Calibration The procedure that follows is applicable to meters that...

Page 8: ...eionized water Dry the probe with a lint free cloth 6 Put the probe in the standard solution with the sensor fully submerged Do not put the probe on the bottom or sides of the beaker 7 Shake the probe...

Page 9: ...when the meter is set to read continuously Air bubbles below the sensor can cause a slow response or error in the measurement Make sure to remove air bubbles before and during measurements 6 2 Sample...

Page 10: ...on relative to the primary ion High concentrations of ions that form insoluble salts with silver will stay on the sensing element and cause measurement errors Strong reducing substances will form a su...

Page 11: ...e bottom or sides of the beaker 7 Shake the probe from side to side to remove air bubbles 8 Stir for 30 to 60 seconds then read the value of the standard solution The meter accepts or rejects the resu...

Page 12: ...junction with deionized water Shake the probe down to remove air bubbles The probe is not conditioned to the sample sufficiently Condition the probe Refer to Preparation for use on page 5 Sample prope...

Page 13: ...f standard solution to the sample Refer to Table 2 Table 2 Standard solution volumes and concentrations Measured sample concentration Volume of standard to add Concentration of standard solution 1 to...

Page 14: ...2370853 Chloride Standard Solution 1000 mg L as Cl 500 mL 18349 Wash bottle polyethylene 500 mL 1 62011 Disposable wipes 11 x 22 cm 280 pkg 2097000 Beaker 30 mL plastic colorless 80 pkg SM5010 Beaker...

Page 15: ......

Page 16: ...932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 rout...

Reviews: