Section 1 Product overview
The Intellical ISECL181 series probes are digital, combination ion-selective electrodes (ISE) that
measure the concentration of chloride in wastewater, drinking water and general water samples. The
probes have a built-in temperature sensor, a non-refillable gel reference and non-replaceable
membranes. Refer to
.
Figure 1 Probe overview
1
Sensing element
3
Sensor protection cap
2
Reference junction
4
Cable
Section 2 Specifications
Specifications are subject to change without notice.
Specifications
Details
Probe type
Digital combination probe with a non-refillable reference junction
and a built-in temperature sensor
Measurement range
0.1 mg/L (3x10
–6
M) to 35,500 mg/L (1 M) Cl
–
Accuracy
±0.02 mV or 0.05% (the larger value)
Sample pH range
pH 1 to 12
Reference type
Ag/AgCl
Reference junction
Double junction (ceramic porous pin and annular porous PTFE)
Slope
58 mV/decade (90 to 110% at 25 °C (77 °F) in linear range per
Nernstian theoretical value)
Linear region
7 mg/L to 35,500 mg/L
Temperature accuracy
±0.3 °C (±0.54 °F)
Temperature sensor type
30 kΩ NTC thermistor
Operating temperature
5 to 50 °C (41 to 122 °F)
Storage temperature
5 to 35 °C (41 to 95 °F)
Response time in linear region < 60 seconds (application dependent)