9705 - Maintenance Procedures
63 of 82
KS
S
- Mar
ch 2
00
6 - E
di
tio
n
1
HIAC
Operator Manual
4) Gently push the brush into the sensor with a twisting motion.
5) Once the cell has been cleaned with the brush, remove the brush and reinstall the
clean-out port cap. Tighten cap snugly.
Do not over-tighten.
6) Flush the sensor with at least 30 cm
3
of clean fluid.
Fig 5-1 :
Insert Brush Into Cleaning Port