3.10.3 Shortening the Adapter Tube Length
The Large Volume Adapter is supplied with a 24-inch long tube. It has a two piece ferrule pre-
installed for connection into the sampling probe. The ferrule can not be re-used. If a shorter
sample tube is desired, then proceed through the following steps.
1) Leave the tube connected to the Large Volume Adapter. Cut the sampling probe end to
the desired length. Make this cut clean and square. Leave the fitting nut on the tube as
shown. Save the tubing that was cut off and observe how the ferrule is installed on it.
2) Carefully observe the loose stainless steel ferrule sleeve that has been supplied. Note
that it has a flat end and the opposite end is sharp. It is a tapered sleeve.
Fig 3-29 :
Ferrule Installation
Fig 3-30 :
Clean Cut of Tube
Summary of Contents for HIAC 9705
Page 1: ...HIAC 9705 Liquid Particle Counting System Operator Manual...
Page 2: ...9705 KSS March 2006 Edition 1 Operator Manual...
Page 6: ...6 of 82 Table of Contents 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Page 10: ...10 of 82 Manual Overview 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Page 22: ...22 of 82 Initialization 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Page 60: ...60 of 82 Operating the 9705 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Page 64: ...64 of 82 Maintenance Procedures 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Page 72: ...72 of 82 Service Procedures 9705 KSS March 2006 Edition 1 Operator Manual HIAC...
Page 78: ......
Page 81: ...9705 User Notes 81 of 82 KSS March 2006 Edition 1 HIAC Operator Manual User Notes...