Hach Programs—Simplified Analysis
Hach Programs—Simplified Analysis
Page 34
5940018HachPrograms.fm
3.5 Hach Program Options
3.5.1
Running a Reagent Blank
Use an actual reading to set the reagent blank, or enter a value manually.
3.5.1.1
Using an Actual Reading
3.5.1.2
Entering a Value Manually
1.
Touch
Options
, then
Reagent Blank
.
2.
If Reagent Blank is set to off, turn it on.
3.
Touch the button on the left that contains the last concentration read.
4.
Enter a reagent blank value and touch
OK
.
5.
Touch
Adjust
to enable the reagent blank. The
Reagent Blank
icon will appear.
Note:
When the reagent blank option is in use, the Reagent Blank icon will appear on the
reading screen.
Note:
The adjustment must be within the limits specified for the program.
In some Hach Programs, the Reagent Blank option allows a reading
or entry of a specific value that should be subtracted from
subsequent measurements. Using a reagent blank shifts the entire
calibration curve along the y-axis, without altering the shape or
slope of the curve. The effect is equivalent to adding a y-intercept
value to the calibration line. In other words:
Concentration = Concentration Factor x Abs – Reagent Blank
1.
Follow the entire Hach procedure (as described in the
DR/2400
Procedures Manual
) substituting deionized water in place of
the sample.
2.
After reading the concentration, touch
Options
, then
Reagent Blank
.
3.
If Reagent Blank is set to off, turn it on.
4.
The concentration will appear in the box on the left of the screen.
5.
Touch
Adjust
to enable the reagent blank.
Summary of Contents for DR2400
Page 2: ...Page 2 5940018Title fm...
Page 6: ...Page 6 5940018TOC fm...
Page 8: ...Page 8 5940018Safety fm...
Page 10: ...Page 10 5940018Spec fm...
Page 12: ...Page 12 5940018InstStop fm...
Page 20: ...Page 20 5940018Getting Started fm...
Page 32: ...Page 32 5940018Operation fm...
Page 46: ...Page 46 5940018StdAdd fm...
Page 48: ...Page 48 5940018SingleWave fm...
Page 50: ...Page 50 5940018MaintStop fm...
Page 56: ...Page 56 5940018Troubleshoot fm...
Page 58: ...Page 58 5940018GenInfo fm...