89
Pour-Thru Cell
A.4 Using the Pour-Thru Cell
The Pour-Thru Cell is intended primarily for those methods that
specifically call for its use. These use liquid reagents, and are
typically denoted as ULR or Rapid Liquid methods. These methods
obtain maximum benefit from the Pour-Thru Cell by avoiding
turbidity from un-dissolved reagents, and provide the precision and
sensitivity required to determine very low concentrations. These
methods require no modification when the Pour-Thru Cell is used.
It is possible to use the Pour-Thru Cell for other methods, with a few
minor modifications. The cell requires a minimum volume of 25 mL
for proper flushing of any previous solution from the cell - so
methods that utilize a 10 mL sample volume cannot be used
directly. Tripling the sample and reagent volumes on these methods
(or use of 25 mL reagents and sample volumes) will provide
enough liquid to use the cell - however, an additional modification is
needed. Since the Pour-Thru Cell has a slightly shorter 1-inch
pathlength than the square cells for which most methods are
calibrated, it is necessary to multiply the results by a pathlength
factor of 0.95, or to perform a standard adjustment (
section 5.4.5 on
page 36
) on any method that utilizes 1-inch square cells. Always
confirm results with a standard solution, to make sure modifications
have been performed correctly.
Although the Pour-Thru cell can be used for the methods described
in
Table 13
, the cell must be thoroughly purged with distilled water
between individual samples.
The Pour-Thru Cell can also be used to run Nitrogen, Ammonia,
Nessler Method and TKN chemistries. Clean the cell by pouring a
few sodium thiosulfate pentahydrate crystals into the cell. Rinse out
the crystals with deionized water.
Important Note: In general, the Pour-Thru Cell cannot be used for
the methods described in
Table 14
.
Table 13 Methods that Require Additional Purging
Aluminum, Aluminon
Chlorine Dioxide, LR
Cobalt, PAN
Copper, Porphyrin
Hardness, Calmagite
Manganese, LR, PAN
Nickel, PAN
Nitrate, MR
Nitrate, HR
Table 14 Methods that Cannot be Used with the Pour-Thru Cell
Aluminum ECR
Arsenic
Barium
Boron, Carmine
Cyanuric Acid
Fluoride
Formaldehyde
Lead, LeadTrak
Mercury
Nickel, Heptoxime
Nitrite, HR
PCB
Phenols
Potassium
Selenium
Silver
Suspended Solids
Sulfate
TPH
Volatile Acids
Zinc
Surfactants, Anionic (Detergents)
Summary of Contents for DR 2800
Page 2: ...Visit us at www hach com...
Page 6: ...Visit us at www hach com...
Page 14: ...14 Installation Figure 5 Installing a Cell Adapter...
Page 18: ...Visit us at www hach com...
Page 46: ...Visit us at www hach com...
Page 78: ...Visit us at www hach com...
Page 80: ...Visit us at www hach com...
Page 82: ...Visit us at www hach com...
Page 86: ...Visit us at www hach com...