
41
Standard Operations
2.
In the result display, press
OPTIONS
>
MORE
>
CHEMICAL FORM
.
3.
A list of available evaluation forms appears. Select the default
setting. Press
OK
to confirm.
4.
Press
EDIT
and then
SAVE
.
The current result and all further measurements will be displayed in
the new chemical form.
5.5.4 Setting Basic Test and Sample Data Settings
1.
Press
OPTIONS
to change test or sample specific settings.
Table 4 Barcode Program Options
Options
Description
Store Off/On
With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.
Abs % Trans
Switches to % transmittance or absorbance readings
Send Data icon
Sends data to a printer, computer, or USB memory stick (Type A)
Timer icon
Functions as a stopwatch. Helps to ensure that the steps of an analysis are correctly timed (e.g.
reaction times, wait times, etc., can be exactly specified). When the specified time has elapsed,
an acoustic signal is emitted. The use of the timer has no influence on the measurement program.
Dilution Factor Off/On
A corrective dilution factor can be entered in order to take account of certain properties.
The number entered at the dilution factor prompt will be multiplied by the result to compensate for
the adjustment. For example, if the sample has been diluted by a factor of 2, enter 2. The default
setting of the dilution factor is 1, corresponding to no dilution.
Note: When a dilution is in effect, the dilution icon will appear on the display.
Note: If you are using undiluted samples, set the dilution factor to 1 again.
Chemical Form
For some stored programs, an alternate chemical form and corresponding text can be selected.
Edit
Modifies an existing program
Save as User Program
Stores the selected parameters as a User Program, see
section 6.1 on page 47
.
Recall Data
Recalls saved measurement data, wavelength scans, or time courses, see
section 5.3 on
page 28
.
Instrument Setup
Basic operation settings of the instrument, see
section 5.2 on page 23
.
Summary of Contents for DR 2800
Page 2: ...Visit us at www hach com...
Page 6: ...Visit us at www hach com...
Page 14: ...14 Installation Figure 5 Installing a Cell Adapter...
Page 18: ...Visit us at www hach com...
Page 46: ...Visit us at www hach com...
Page 78: ...Visit us at www hach com...
Page 80: ...Visit us at www hach com...
Page 82: ...Visit us at www hach com...
Page 86: ...Visit us at www hach com...