58
Advanced Operations
6.2.1 Using the Standard Addition Option
1.
From the Main Menu, select
STORED PROGRAMS
. Select the
required programs.
2.
Press
START
.
3.
Analyze a sample without added standard solution in
accordance with the instructions in the Procedures Manual.
When the measurement is complete, leave the sample cell in
the cell holder.
4.
Press
OPTIONS
>
MORE
>
STANDARD ADDITION
.
An overview of the data of the standard addition procedure is
displayed.
5.
Press
OK
to accept the standard values for standard
concentration, sample volume (total) and standard addition
volume. Press
EDIT
to change any of these values.
6.
Press the key of the value you want to change. Use the
alphanumeric keypad to change the value and press
OK>NEXT
.