Table 2 Acceptable and unacceptable cleaning solutions
Acceptable
Do not use
Dish detergent and water
Concentrated bleach
Window cleaner
Kerosene
Isopropyl alcohol
Gasoline
Dilute acids
Aromatic hydrocarbons
Replace the desiccant
N O T I C E
Do not operate the sensor without desiccant beads or with green desiccant beads. Permanent damage to the
sensor can occur.
Immediately replace the desiccant when it changes to green. Refer to
.
Note: It is not necessary to remove the desiccant container from the desiccant hub to install new desiccant.
, make sure that the O-ring is clean and has no dirt or debris. Examine the O-
ring for cracking, pits or sign of damage. Replace the O-ring if it has damage. Apply grease to dry or
new O-rings to make installation easier, to get a better seal and to increase the life of the O-ring.
For the best performance, make sure to install the desiccant container vertically with the end cap
pointed down. Refer to
on page 8.
Note: When the beads just begin to turn green, it may be possible to rejuvenate them by heating. Remove the
beads from the canister and heat them at 100-180 ºC (212-350 ºF) until they turn orange. Do not heat the canister.
If the beads do not turn orange, they must be replaced with new desiccant.
Figure 11 Replace the desiccant
English
17
Summary of Contents for AV9000
Page 2: ...English 3 Fran ais 22 Espa ol 42 Portugu s 63 83 100 119 2...
Page 39: ...Fran ais 39...
Page 59: ...Espa ol 59...
Page 62: ...45 1 14 m 1 1 1 2 48 1 21 m 1 0 2 2 62 Espa ol...
Page 80: ...80 Portugu s...
Page 85: ...ESD 1993 4 15 OSHA CFR 1910 146 Permit Required Confined Spaces 250000 19 5 23 5 H2S 10 ppm 85...
Page 89: ...4 FL1500 1 AV9000S 3 2 89...
Page 91: ...1 InSight 2 InSight 3 4 5 6 1 6 2 6 3 8 93 6 1 3 2 4 2 1 7 1 2 91...
Page 92: ...3 8 7 1 2 3 1 B AV9000 5 fps AV9000 AV FL900 92...
Page 95: ...2 1 2 10 3 10 4 5 6 10 1 2 3 2 95...
Page 96: ...11 11 5 O O O O O 88 100 180 C 212 350 F 11 4 88 96...
Page 97: ...in 97...
Page 103: ...AV Sigma FL AS950 1 AV AV9000 FL AS950 AV9000 117 AV Sigma AV9000 1 1 6 2 7 AV 3 8 4 9 5 x 103...
Page 106: ...3 FL900 1 4 FL1500 1 AV9000S 3 2 106...
Page 109: ...6 1 3 2 4 2 1 7 1 2 3 8 109...
Page 110: ...7 1 2 3 1 B AV9000 5 fps AV9000 AV FL900 Sensor Port Set Up Reversed Sensor 110...
Page 113: ...2 1 2 10 3 10 4 5 6 113...
Page 114: ...10 1 2 3 2 11 11 5 O O O O O O 105 100 180 C 114...
Page 115: ...11 4 105 115...
Page 116: ...116...
Page 121: ...ESD XXX 1993 4 15 OSHA CFR 1910 146 250 000 19 5 23 5 H2S 10 ppm 121...
Page 122: ...AV Sigma FL AS950 1 AV AV9000 FL AS950 AV9000 134 AV Sigma AV9000 1 1 6 2 7 AV 3 8 4 9 5 x 122...
Page 125: ...3 FL900 1 4 FL1500 1 AV9000S 3 2 125...
Page 126: ...5 AS950 1 FL FL900 FSDATA FSDATA FSDATA FL1500 FL1500 FL1500 FSDATA Sigma 910 950 126...
Page 128: ...1 7 1 2 3 8 7 1 2 3 1 AV9000 5 fps AV9000 AV FL900 128...
Page 130: ...FL1500 AS950 FL1500 FL1500 FSDATA Sigma Sigma 130 Sigma 1 PC Insight 2 9 3 2 9 9 4 9 1 130...
Page 131: ...2 1 2 10 3 10 4 5 6 10 1 2 3 2 131...
Page 132: ...2 11 11 5 O O O 124 100 180 C 212 350 F 11 132...
Page 133: ...4 124 133...
Page 134: ...134 1 5 8755500 8542000 3390 O 1 176 ID x 0 070 OD 5252 100 50mL 2 7724700 134...
Page 137: ......