配置继电器

如果使用控制器的继电器接头,需要配置继电器。控制器的四个继电器通
过所测浊度或诊断实现通电或断电(开或关)。

1.

 

MENU

 

并选择

 CONFIGURE>SET RELAY

A

B

 D

)。

2.

选择一个配置选项。

选项

说明

SET FUNCTION

(设定功能)

将继电器设为控制、报警或状态继电器。

ALARM

(报警)

(默认)

— 

该继电器具有单独的报警上下限和死区,响应所

测浊度进行操作。

CONTROL

(控制)

 — 

该继电器具有状

态、设定值、死区及过载定时器,响应所测浊度进行操作。

STATUS

(状态)

 — 

该继电器响应所选的系统诊断状况进

行操作。

SET TRANSFER

(设定转换)

设定继电器的转换状态

 — 

断电(默认)或通电。校准期

间,继电器可以保持或转换状态,或保持启用。选择转换
时,继电器将保持在转换状态。

ACTIVATION

(激

活)

该菜单随所选继电器功能(报警继电器、控制继电器或状
态继电器)而变。请参阅

设置功能

每种继电器都有一个介绍表格。请参阅适用表格了解配置
选项。

报警继电器

选项

说明

LOW ALARM

(低位警

报)

设定报警下限

 — 0.00

(默认)至

 100.00

。所测浊度

低于报警下限时继电器通电。

HIGH ALARM

(高位警

报)

设定报警上限

 — 0.00 

 100.00

(默认)。所测浊

度高于报警上限时继电器通电。

LOW DEADBAND

(低

位死区)

设定所测浊度增加到高于报警下限之后使继电器保
持接通的浊度范围

 — 0.00

(默认)至

 10.00

HIGH DEADBAND

(高

位死区)

设定所测浊度值降到低于报警上限之后使继电器保
持接通的浊度范围

 — 0.00

(默认)至

 10.00

OFF DELAY

(关闭延时) 设定继电器的关闭延时

 — 0

(默认)至

 300 

秒。

ON DELAY

(打开延时) 设定继电器的打开延时

 — 0

(默认)至

 300 

秒。

控制继电器

选项

说明

PHASE

(状态)

设定继电器处于高位或低位。高位(默认)

— 

所测浊度

值高于设定值时继电器通电。低位

 — 

所测浊度值低于

设定值时继电器通电。

SET SETPOINT

(设

定点)

设定使继电器通电的设定值(所测浊度)

— 0.00 

100.00

(默认)。

DEADBAND

(死区) 设定使继电器保持接通的所测浊度范围

 — 0.00

(默认)

 10.00

。对于高位继电器,是指所测浊度值降至低于设

定值之后的浊度范围。对于低位继电器,是指所测浊度
值升至高于设定值之后的浊度范围。

OVERFEED TIMER

(过载定时器)

设定继电器保持接通的限制时间

 — 0.0

(默认)至

999.9 

分钟。有关过载定时器操作的详情,请参阅

控制继

电器的过载定时器

 

 43

OFF DELAY

(关闭

延时)

设定继电器的关闭延时

 — 0

(默认)至

 300 

秒。

ON DELAY

(打开延

时)

设定继电器的打开延时

 — 0

(默认)至

 300 

秒。

状态继电器

选项

说明

SET STATUS
TYPE

(设定状

态类型)

设定继电器的状态类型。

ALL

(所有)(默认)

— 

出现

流动槽未知

流动槽变脏

状况时,继电器通电。

WARN

(警告)

— 

出现

流动槽变脏

状况时,继电器通电。

FAIL

(故障)

— 

出现

故障

流动槽未知

状况时,继电器通

电。请参阅

故障排除

 

 47

了解状况列表。

3.

 

ENTER

 

确认。

控制继电器的过载定时器
过载定时器设置用于防止所测浊度高于设定值或死区设置(比如因为传感
器损坏或工艺扰动)的情况使控制继电器保持连续接通。过载定时器设置
可限制继电器及其所连控制元件保持接通的时间,不受其他情况影响。

如果使用继电器过载定时器并且出现超时,继电器的状态指示灯将在测量
屏幕上闪烁,直到重置过载定时器为止。按

 

Menu

(菜单)并选择

中文

   

43

Summary of Contents for Accu4

Page 1: ...DOC023 97 80356 Accu4 T53 8320 01 2013 Edition 1 User Manual...

Page 2: ...English 3 29 2...

Page 3: ...bubble trap on page 27 Mechanical Light sources Two near infrared 880 nm wavelength LEDs Specification Details Power requirements 5 V supplied by the controller to the light sources and detectors Proc...

Page 4: ...1 of span per C 1 Contact the manufacturer if a longer distance is necessary General information In no event will the manufacturer be liable for direct indirect special incidental or consequential dam...

Page 5: ...f a fuse or current limiting device This symbol indicates that the marked item requires a protective earth connection If the instrument is not supplied with a ground plug on a cord make the protective...

Page 6: ...e by any applicable regulation and to have sufficient and appropriate mechanisms in place for compliance with applicable regulations in the event of malfunction of the instrument The Accu4 low range t...

Page 7: ...0 mm 4x 4 Mounting bracket 9 Pan head screws M6 x 1 0 x 100 mm for pipe mount 4x 5 Ground screw green 10 Pan head screws M6 x 1 0 x 150 mm for panel mount 4x Installation W A R N I N G Multiple hazard...

Page 8: ...er source Protected from corrosive fluids Note If exposed to direct sunlight the operating temperature of the controller may increase above the specified limit Direct sunlight may also decrease the di...

Page 9: ...hold the cables securely when tightened Do not put more than one cable in a cable strain relief Close all unused cable entry holes with applicable plugs to keep out moisture Close all unused cable st...

Page 10: ...g strip General wiring procedure W A R N I N G Potential Electrocution Hazard Always disconnect power to the instrument when making electrical connections Connect wiring to the controller as shown in...

Page 11: ...e controller with conduit or a power cable Make sure that a circuit breaker with sufficient current capacity is installed in the power line The circuit breaker size is based on the wire gauge used for...

Page 12: ...oise may interfere with the sensor signal Table 1 AC power wiring information Input voltage Terminal Description Color North America Color EU 115 VAC single phase 3 Hot Black Brown 230 VAC single phas...

Page 13: ...put 1 and Outputs 2 that represent the measured turbidity Use the analog outputs for analog signaling or to control external devices Refer to Figure 5 on page 10 and General wiring procedure on page 1...

Page 14: ...B C and D that are opened or closed by the measured turbidity or diagnostics Refer to Configure the relays on page 19 The relay outputs are not powered Refer to General wiring procedure on page 10 Fi...

Page 15: ...TTL terminals When this connection is broken the applied hold or transfer releases Note The TTL input feature is disabled during calibration and system diagnostic tests that are started by the user Pl...

Page 16: ...tion ball valve in IPS 2 Isolation ball valve in IPS 5 Process input 3 To open drain 6 Inlet in NPT female Connect the sample stream Install the sample line into a larger process pipe to minimize inte...

Page 17: ...U Go to the Main Menu screen and cancel the procedure to change a setting ESC Go back one menu level or cancel the procedure to change a setting ENTER Confirm a menu selection or confirm changes to a...

Page 18: ...2 value mA Startup Select the language Select the display language default English 1 Push MENU and select CONFIGURE LANGUAGE 2 Use the UP and DOWN arrows to select the language 3 Push ENTER Start the...

Page 19: ...d stay at 20 mA SET TRANSFER Set the transfer value of the analog output 4 or 0 to 20 mA default 20 mA During calibration the analog outputs can be held transferred or kept active When transfer is sel...

Page 20: ...low HIGH default The relay is energized when the measured turbidity is higher than the setpoint LOW The relay is energized when the measured turbidity is lower than the setpoint SET SETPOINT Set the...

Page 21: ...alibration only W A R N I N G Chemical exposure hazard Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled Refer to the cu...

Page 22: ...Pour the sample of known value or prepared standard suspension 20 NTU or 40 NTU into the calibration funnel until the flow chamber is full 9 Put the calibration funnel in the funnel holder Enter the v...

Page 23: ...ory defaults Enable or disable the passcode Enable the passcode to prevent unauthorized changes to the configuration settings and calibrations The passcode is disabled by default The passcode is 3456...

Page 24: ...7 for the replacement fuse kit 3 Close the controller Turn the four screws to tighten Replace a light source or detector D A N G E R Electrocution hazard Remove all power from the instrument and relay...

Page 25: ...material Clean the flow chamber Refer to Clean the flow chamber on page 23 Table 5 Status messages continued Message Description Solution CHAMBER UNKNOWN The sensor cable wiring is not connected or n...

Page 26: ...y meter on the NO and NC relay output terminals to confirm the relay status selected SIM SENSOR Set the simulated turbidity measurement value 0 00 to 100 00 default 0 35 NTU Use this option to confirm...

Page 27: ...d Use of non approved parts may cause personal injury damage to the instrument or equipment malfunction The replacement parts in this section are approved by the manufacturer Note Product and Article...

Page 28: ...t source or detector 1 8220F1078 Accessories Description Quantity Item no Model 8320 turbidity sensor Bubble trap head regulator 1 4668000 Lubricant silicone 2 g 6M0920 Flask class A 2 L 1 2489454 For...

Page 29: ...7 psig 3 8 LPM 1 GPM 9 5 0 19 1 9 LPM 0 05 0 5 GPM 1 9 LPM 0 5 GPM 48 880 nm LED 5 V in NPT 3 8 in in NPT PVC PPO x x 285 x 172 x 400 mm 11 21 x 6 79 x 15 75 in 4 5 kg 10 lb 15 T53 LCD 128 x 64 LED 13...

Page 30: ...00 20 00 mA 4 00 20 00 mA 0 004 mA 12 600 NEMA 4X 13 mm in 2 3 kg 5 lb T 80 mA 250 V 230 V 5 mm x 20 mm T 100 mA 250 V 115 V 5 mm x 20 mm 15 Accu4 0 000 100 0 NTU 1 000 NTU 10 00 NTU NTU TEF FNU FTU...

Page 31: ...pment Regulation IECS 003 A A Cet appareil num rique de classe A r pond toutes les exigences de la r glementation canadienne sur les quipements provoquant des interf rences FCC 15 A FCC 15 1 2 FCC 15...

Page 32: ...1 8320 1 6 2 2 7 3 2 8 in NPT 4 9 5 in NPT 10 2 3 2 1 3 4 2 4 32...

Page 33: ...3 1 6 inID 8 2 7 M6 8 3 2 8 M6 x 1 0 x 20 mm 4 4 9 M6 x 1 0 x 100 mm 4 5 10 M6 x 1 0 x 150 mm 4 9 m 30 ft 29 1000G3088 001 4 33...

Page 34: ...4 NEMA IP NEMA 4X IP66 3H1091 3H1230 4 3 11 4 mm 0 17 0 45 in ESD ESD 5 34...

Page 35: ...5 1 2 5 2 RS232 6 2 3 TTL 7 4 4 8 6 NEMA IP 6 35...

Page 36: ...3 m 10 ft 18 12 AWG 3 m 10 ft 60 C 140 F 18 AWG 1 5 35 35 3 4 1 115 VAC 3 230 VAC 4 115 230 VAC 2 N PE 35 7 2 36...

Page 37: ...7 2 TB 1 12 18 13 19 14 20 15 21 16 22 0 00 20 00 mA 4 00 20 00 mA 1 2 5 35 35 42 600 37...

Page 38: ...A B C D 43 35 8 3 18 12 AWG 18 AWG 5 A 30 V RMS 42 2 V PEAK 60 VDC 30 V RMS 42 2 V PEAK 60 VDC 8 1 A 5 2 C 6 3 0 1 F 600 V 7 100 230 VAC 4 100 1 W 8 2 N 3 NC COM TTL 38...

Page 39: ...TTL 5 35 TTL HOLD XFER ACTIVE HOLD XFER 42 43 ACTIVE TTL TTL TTL TTL 9 10 9 1 in NPT 6 2 7 in IPS 3 in IPS 8 in NPT 4 9 in IPS 5 in IPS 39...

Page 40: ...10 1 in NPT 4 in IPS 2 in IPS 5 3 6 in NPT 11 11 1 2 12 T53 4 40...

Page 41: ...12 1 2 4 MENU Esc ENTER RIGHT LEFT UP DOWN 13 MENU ESC 30 RIGHT LEFT ENTER UP DOWN 13 1 1 3 1 mA 3 2 2 1 A C RELAY A C 2 3 DOWN 2 mA 1 MENU CONFIGURE LANGUAGE 2 UP DOWN 3 ENTER 41...

Page 42: ...N OFF ACCU4 8 UP DOWN RIGHT 3 ENTER 1 MENU CONFIGURE SET OUTPUT 1 2 2 SET 4mA or 0mA VALUE 4mA 0mA 4 mA 0 mA 0 00 100 00 SET 20mA VALUE 20 mA 20 mA 0 00 100 00 0 4 mA 20 mA 20 mA SET TRANSFER 4 0 20 m...

Page 43: ...ATION LOW ALARM 0 00 100 00 HIGH ALARM 0 00 100 00 LOW DEADBAND 0 00 10 00 HIGH DEADBAND 0 00 10 00 OFF DELAY 0 300 ON DELAY 0 300 PHASE SET SETPOINT 0 00 100 00 DEADBAND 0 00 10 00 OVERFEED TIMER 0 0...

Page 44: ...4000 NTU Formazin MSDS SDS 4 NH2 2H2SO4 5 CH2 6N4 50 1 0 2 2 1000 mL 3 1 5 00 500 mL 4 2 50 00 500 mL 5 50 0 mL 1 50 0 mL 2 4000 NTU 6 25 C 3 C 77 F 5 4 F 24 7 1 8 20 NTU 40 NTU a 20 NTU 10 mL 4000 N...

Page 45: ...AL ENTER PRIMARY or SAMPLE CAL CONFIRM CAL OK CONFIRM FAILURE ENTER CAL EXIT CAL REPEAT 4 ENTER Confirm Active ESC 1 2 3 42 4 ENTER CONFIRM ACTIVE Confirm Active 5 ENTER 2 mA 0 20 mA 4 mA 20 mA 0 mA 1...

Page 46: ...3456 1 MENU CONFIGURE SET PASSCODE 2 UP DOWN TEST MAINT CALIBRATE TEST MAINT 3 ENTER 1 a 3 CALIBRATE b c 2 1 8 3 4 5 6 7 O 8 1 8 1 35 2 5 35 49 3 O 44 46...

Page 47: ...TER 5 5 ANALYZER OK ANALYZER FAIL Source 1 or 2 OK 1 2 SOURCE 1 or 2 FAIL 1 2 46 DETECT 1 or 2 OK 1 2 DETECT 1 or 2 FAIL 1 2 46 5 CHAMBER OK CHAMBER DIRTY 46 CHAMBER UNKNOWN 46 RLY A B C D ACTIVE A B...

Page 48: ...VERFEED RESET RLY A B C D COUNTING A B C D 1 MENU TEST MAINT 2 CHECK CAL HOLD 30 OUTPUT 1 or 2 1 2 0 4 20 mA 4 05 mA RELAY A B C D A B C D SIM SENSOR SIM 0 00 100 00 0 35 NTU 3 ENTER ESC 1 29 2 3 4 5...

Page 49: ...x 1 0 x 100 mm 4 8H1217 400 M6 x 1 0 x 150 mm 4 8H1217 600 1 1000A4F3249 101 8320 1 8220G1060 O 1 8220G1330 204 1 8220C2F1076 1 8320A1074 001 O 1 8220G1320 204 O 1 5H1288 O 1 5H1251 1 8220F1078 8320...

Page 50: ......

Page 51: ......

Page 52: ...ach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE S rl 6 route de Com...

Reviews: