Table of contents
Specifications
on page 3
Operation
on page 18
General information
on page 4
Maintenance
on page 23
Installation
on page 7
Troubleshooting
on page 24
User interface and navigation
on page 17
Replacement parts and accessories
on page 27
Startup
on page 18
Specifications
Specifications are subject to change without notice.
Model 8320 turbidity sensor
Specification
Details
Operational:
Flow rate
0.19 to 26.5 LPM (0.05 to 7 GPM)
Ambient conditions
0 to 60 °C (32 to 140 °F)
Sample temperature
range
0 to 60 °C (32 to 140 °F)
Pressure range
0–3.4 bar at 20 °C (0–50 psig at 68 °F)
Pressure drop
0.0001 bar at 0.36 LPM (0.0017 psig at 0.1 GPM)
Residence time
3.8 LPM (9.5 seconds at 1 GPM)
Air venting
Integral bubble trap for 0.19 to 1.9 LPM (0.05 to
0.5 GPM) flows.
Installation of an external bubble trap and a restrictor
valve on the sensor outlet is recommended for flows
above 1.9 LPM (0.5 GPM) with air in the sample.
Refer to
Use an external bubble trap
on page 27.
Mechanical:
Light sources
Two near-infrared (880 nm wavelength) LEDs
Specification
Details
Power requirements
5 V supplied by the controller to the light sources
and detectors
Process connections
½–in. NPT female standard; adaptable to
3
/
8
–in. or
¼–in. NPT, barb or tube fittings
Wetted materials
PVC, polycarbonate, polystyrene, PPO, nitrile and
Buna-N
Enclosure
Molded fiber-glass-reinforced polyester (flame
retardant) with four brackets for surface mounting
Dimensions (W x D x H)
285 x 172 x 400 mm (11.21 x 6.79 x 15.75 in.)
Net weight
4.5 kg (10 lb) approximately
Certifications
15 year environmental friendly use period
Model T53 controller
Specification
Details
Operational:
Display
Graphic dot matrix LCD, 128 x 64 pixels with LED
backlighting; 13 mm (½–in.) main character height;
3 mm (
1
/
8
–in.) auxiliary information character height;
menu screens contain up to six lines of text
Ambient conditions
Operation: –20 to 60 °C (–4 to 140 °F); 0–95% relative
humidity, non-condensing
Storage: –30 to 70 °C (–22 to 158 °F); 0–95% relative
humidity, non-condensing
Pollution degree
2
Installation category
II
Protection class
I
Altitude
2000 m (6561 ft) maximum
English
3
Summary of Contents for Accu4
Page 1: ...DOC023 97 80356 Accu4 T53 8320 01 2013 Edition 1 User Manual...
Page 2: ...English 3 29 2...
Page 32: ...1 8320 1 6 2 2 7 3 2 8 in NPT 4 9 5 in NPT 10 2 3 2 1 3 4 2 4 32...
Page 34: ...4 NEMA IP NEMA 4X IP66 3H1091 3H1230 4 3 11 4 mm 0 17 0 45 in ESD ESD 5 34...
Page 35: ...5 1 2 5 2 RS232 6 2 3 TTL 7 4 4 8 6 NEMA IP 6 35...
Page 37: ...7 2 TB 1 12 18 13 19 14 20 15 21 16 22 0 00 20 00 mA 4 00 20 00 mA 1 2 5 35 35 42 600 37...
Page 40: ...10 1 in NPT 4 in IPS 2 in IPS 5 3 6 in NPT 11 11 1 2 12 T53 4 40...
Page 50: ......
Page 51: ......