• Reminders—wrench symbol shows on the display and yellow status indicator light. The time for a
maintenance task has passed.
1.
Push
diag
to access the DIAG/TEST menu.
2.
Select an option.
Option
Description
DIAGNOSTICS
Shows the errors and warnings that currently are on the instrument or on the installed
modules. The analyzer is in operation with the active warnings or reminders until they
are acknowledged or reset. Then, the display background will return to white.
PROGNOSYS
Shows the variables which trigger the service indicator and the measurement health
indicator on the display.
CURRENT STATUS
Shows the current instrument statuses that follow: OPERATION—Current
measurement mode. SAMPLE CHANNEL—Current sample channel. STEP STATUS
—Current step in the measurement cycle. STEP TIME—Step time remaining.
MINUTES LEFT—Minutes left in current step. COMPLETION—% completed of the
measurement cycle.
ANALYZER HELP
Shows all possible errors, warnings and reminders with troubleshooting hints.
PERFORM TEST
Examines individual parts of the analyzer. Refer to
for more details about the individual test options.
OUTPUTS
Shows the current status of the 4–20 mA and relay outputs with the options to
examine, hold and simulate the outputs. Refer to
on page 13 for more
information.
VIEW LED
Illuminates the colorimeter cell for improved viewing during troubleshooting. The cell
can be illuminated from 1 to 999 seconds.
MODBUS STATS
Shows the status of the Modbus ports: sensor, controller, network and service. Shows
the number of good and bad transmissions.
SERVICE
Shows the service parts information and the history. SERVICE PART—shows the last
and the next service date and the remaining days. PART INFORMATION—Shows the
replaced part and the current run time. UPCOMING SERVICE—Shows the next part
which needs to be replaced. SERVICE HISTORY—Shows the date and time of the
replaced parts.
SYSTEM DATA
Shows the system information. TEMPERATURE—Shows the measured temperature
of the A/D device in Celsius (C). POWER SOURCE FREQUENCY—Shows the line
power frequency (Hz). POWER SOURCE VOLTAGE—Shows the line power voltage
(V). 12 V VOLTAGE—Shows the measured power supply voltage (V DC). 3.3 V
VOLTAGE—Shows the measured regulated 3.3 V supply (V DC). 12 V CURRENT—
Shows the measured 12 V power supply current (Amps).
I2C DATA
Shows the display information (I
2
C) and the version number.
OVERFEED RESET
Resets the overfeed timer.
Output options
The output menu shows the current status of the 4–20 mA and relay outputs with the options to
examine, hold and simulate the outputs.
1.
Push
diag
and select OUTPUTS.
2.
Select an option.
Option
Description
TEST 4–20 mA
Examines the 4–20 mA outputs from 1–4.
TEST RELAY
Examines the relays A–D. Sets the relays to on or off.
English
13
Summary of Contents for 5500sc
Page 2: ...English 3 Fran ais 18 Espa ol 35 Portugu s 52 69 84 100 116 2...
Page 6: ...Figure 1 Funnel and colorimeter access Figure 2 Clean the sample cell and stir bar 6 English...
Page 69: ...69 73 70 76 70 80 1 1 30 60 90 365 70 X 71 X 73 X1 X2 73 X3 Y X X X X X X 1 10 2 15 3 69...
Page 71: ...MSDS SDS 1 1 2 2 1 4 1N 5 MSDS SDS 70 1 2 1 71...
Page 72: ...2 3 3 72...
Page 74: ...4 5 1 4 2 2 3 2 74...
Page 76: ...77 77 77 77 77 R1 R1 77 77 R1 R1 R2 77 77 R2 76...
Page 83: ...1 5500sc 2L 6776102 2 5500sc 2L 6776202 3 5500sc 2L 6776302 1 5500sc 2L 6776402 83...
Page 84: ...84 88 85 91 85 96 1 1 30 60 90 365 85 X 86 X 88 X1 X2 88 X3 Y X X X X X X 1 10 2 15 3 1 1 84...
Page 86: ...1 2 MSDS SDS 1 1 2 2 1N 5 1 4 MSDS SDS 85 1 2 86...
Page 87: ...1 2 87...
Page 90: ...5 1 x 4 2 x 2 3 x 2 MSDS SDS 1 85 2 3 4 5 2 6 7 2 8 9 10 SD SD 90...
Page 102: ...1 2 MSDS SDS 1 1 2 2 1N 5 1 4 MSDS SDS 101 1 2 102...
Page 103: ...1 2 103...
Page 106: ...5 1 4x 2 2x 3 2x MSDS SDS 1 101 2 3 4 5 2 6 7 2 8 9 10 SD SD 106...
Page 118: ...1 2 MSDS SDS 1 2 1N 5 MSDS SDS 117 1 2 118...
Page 119: ...1 2 119...
Page 122: ...5 1 4x 2 2x 3 2x MSDS SDS 1 117 2 3 4 5 6 7 8 9 10 SD SD 122...
Page 131: ......