Option
Description
COLORIMETER
HEATER
Set the colorimeter heater settings from 20–60 °C (68–140 °F). The measured
value is shown.
SAMPLE HEATER
Set the sample heater settings from 20–60 °C (68–140 °F). The measured value
is shown.
COLORIMETER
Start an automatic test which increases the optical LED duty cycle in 5%
increments. This starts from 0% until the output reaches the saturation. The A2D
counts are shown for 0%, then % before the saturation and the first saturation
value (%).
STATUS LED
Examine the front panel status LED indicator. The test cycles continuously until
interrupted: off, red, green, yellow.
A2D
Set the colorimeter LED intensity to examine the cell transmittance for the A2D
output.
AIR PUMP
Change and control the air pressure. SET SETPOINT—Range: 1–9.99 psi. LOW
and HIGH DEADBAND—Range: 0–1 psi. SET LOW and HIGH VALUE—Range:
5–99.99 psi. START—Start the air pump with the entered settings.
FAN
Set to adjust the fan duty cycle.
ANALYZER TYPE
For use by manufacturer technical support only.
SELECT SCRIPT
Toggle between normal instrument script and test script.
SET CHANNELS
For use by manufacturer technical support only.
Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Item no.
Air pump filter
2718
Bottle, reagent, 2 liter
9395000
Capillary assembly, silica
6786900
Capillary assembly, HR and LR phosphate
6786902
Cell cap
6767800
Cell shroud
6773100
Colorimeter assembly, silica
6786800
Colorimeter assembly, LR phosphate
6786801
Colorimeter assembly, HR phosphate
6786802
Colorimeter cell
6768000
Colorimeter cover
6766900
Fan assembly
6789800
Fan filter plug
6789300
English
15
Summary of Contents for 5500sc
Page 2: ...English 3 Fran ais 18 Espa ol 35 Portugu s 52 69 84 100 116 2...
Page 6: ...Figure 1 Funnel and colorimeter access Figure 2 Clean the sample cell and stir bar 6 English...
Page 69: ...69 73 70 76 70 80 1 1 30 60 90 365 70 X 71 X 73 X1 X2 73 X3 Y X X X X X X 1 10 2 15 3 69...
Page 71: ...MSDS SDS 1 1 2 2 1 4 1N 5 MSDS SDS 70 1 2 1 71...
Page 72: ...2 3 3 72...
Page 74: ...4 5 1 4 2 2 3 2 74...
Page 76: ...77 77 77 77 77 R1 R1 77 77 R1 R1 R2 77 77 R2 76...
Page 83: ...1 5500sc 2L 6776102 2 5500sc 2L 6776202 3 5500sc 2L 6776302 1 5500sc 2L 6776402 83...
Page 84: ...84 88 85 91 85 96 1 1 30 60 90 365 85 X 86 X 88 X1 X2 88 X3 Y X X X X X X 1 10 2 15 3 1 1 84...
Page 86: ...1 2 MSDS SDS 1 1 2 2 1N 5 1 4 MSDS SDS 85 1 2 86...
Page 87: ...1 2 87...
Page 90: ...5 1 x 4 2 x 2 3 x 2 MSDS SDS 1 85 2 3 4 5 2 6 7 2 8 9 10 SD SD 90...
Page 102: ...1 2 MSDS SDS 1 1 2 2 1N 5 1 4 MSDS SDS 101 1 2 102...
Page 103: ...1 2 103...
Page 106: ...5 1 4x 2 2x 3 2x MSDS SDS 1 101 2 3 4 5 2 6 7 2 8 9 10 SD SD 106...
Page 118: ...1 2 MSDS SDS 1 2 1N 5 MSDS SDS 117 1 2 118...
Page 119: ...1 2 119...
Page 122: ...5 1 4x 2 2x 3 2x MSDS SDS 1 117 2 3 4 5 6 7 8 9 10 SD SD 122...
Page 131: ......