Section 1 Specifications
Specifications are subject to change without notice.
General specifications
Specification
Details
Dimensions (W x D x H)
42 x 31.5 x 18 cm (16.5 x 12.5 x 7 in.)
Enclosure rating
IP62
Weight
11.3 kg (25 lbs)
Installation environment
Indoor
Mount
Wall
Pollution degree/installation
category
2/II
Protection class
I
Altitude
Maximum 2000 m (6562 ft)
Light source
Class 1 LED (light emitting diode) with peak wavelength of
610 nm. Estimated 50,000 hours minimum life; hard water
and soft water indicator.
Detector
Silicon photo detector
Optical path length
8.89 mm (0.35 in.)
Time delay
One reading above the trip point is necessary to trigger a
soft-to-hard water alarm transition. One reading below the
trip point is necessary to trigger a hard-to-soft water alarm
transition.
Trip point
0.3, 1, 2, 5, 10, 20, 50 and 100 mg/L. Refer to
Reagents
Buffer and indicator solution. 500 mL of each reagent every
2 months.
Reagent shelf life
1 year
Power requirements
115/230 VAC, 70 VA, 50/60 Hz, 1.25 A Fuse
Electrical connections
Two 3-wire barrier terminal blocks. Wire range: 18–12 AWG
(0.75–1.0 mm
2
)
Hard water alarm relay
SPDT relay, set to on when the hard water indicator is on.
The alarm can be disabled.
Contact rating
5 A resistive at 100–240 VAC
Operating temperature
5 to 40 °C (32 to 104 °F)
Storage temperature
-40 to 60 °C (-40 to 140 °F)
Operating humidity
Relative humidity: 5–95% at different temperatures, non-
condensing
Sampling rate
New sample: every 1.9 minutes ±5% at 60 Hz; 2.3 minutes
±5% at 50 Hz
English
3
Summary of Contents for 5410001
Page 1: ...DOC023 53 80422 SP 510 03 2022 Edition 13 User Manual...
Page 2: ......
Page 11: ...Figure 3 Dimensions for wall installation Figure 4 Door clearance English 9...
Page 28: ......