Option
Description
Zero calibration
bottle
Make sure this is disabled by unchecking the box as it is not relevant for this sensor.
Stop parameters
If this button is pressed you can view or change the existing values or restore the
default values. It is
highly recommended
to leave these parameters at their default
values. These values apply to manual calibrations with the
Auto-End
parameter
enabled.
Configure manual calibration
1.
Set the parameters for a manual sensor calibration:
Option
Description
Calibration mode
Select
Zero calibration
or
High level adjustment
. If using a high range sensor you also
have the option of
100% humid air calibration
.
Note: If zero calibration or 100% humid air calibration is selected, no other parameters
are required. The following are only required for high level adjustment.
Cal. sample
Set to
In line sample
,
Gas bottle
or
Factory parameters
. If factory parameters is
selected, the Ksv value is displayed but can be changed.
These additional parameters are required if in line sample or gas bottle has been
selected as the calibration sample:
Medium
This is automatically set to
Liquid
if in line sample has been selected as the calibration
sample, or
Gas
if gas bottle has been selected.
Gas unit type
Either
Partial
or
Dissolved
are available for an in line sample. If gas bottle was selected
this is set to
Fraction
.
Gas unit
The list of available units depends on unit type selected above.
Liquid
Select Water for the for the K1100 sensor (default) or Beer for the M1100 sensor.
Reference value
Enter the reference value for calibration.
Zero calibration
With this method, the sensor should be removed from the sample and exposed to pure N
2
gas. It is
recommended to use the specially designed portable calibration device for this purpose.
Press
Start
to start the calibration.
A screen is displayed showing the measured values and length of time the sensor has been under
calibration. These values are continually refreshed.
The value
% last calibration
is an informational message showing the difference between the
current and previous sensor calibrations.
The
Signal within range
and
Stability reached
boxes indicate whether the calibration is within
acceptable limits. When both boxes indicate
YES
, press
Finish
to accept the new calibration. If one
or both boxes continue to show
NO
you can still perform a calibration but it is
not recommended
and the calibration should be aborted by pressing the
Cancel
button.
In the case of a calibration failure, attempt a second calibration after about 5 minutes. If the second
attempt also fails, then refer to your Hach representative for advice.
Note: If the Auto-End parameter is enabled then the calibration will be considered successful when the parameters
defined in Stop parameters are met.
If you have not accepted or cancelled the calibration after an elapsed time of 10 minutes then the
process will time-out.
100% humid air calibration (high range sensors only)
With this method, the sensor should be removed from the sample and exposed to air saturated with
humidity. Do this by putting a drop of water in the calibration cap before installing the cap on the
sensor. Press
Start
to start the calibration. The process is then the same as for the
Zero calibration
described previously.
English
19
Summary of Contents for 51 Series
Page 2: ...English 3 Deutsch 22 Fran ais 43 Espa ol 63 e tina 84 Svenska 104 124 Magyar 145 T rk e 165 2...
Page 127: ...ESD 127...
Page 128: ...2000 m 6562 ft 2000 m GFCI GFI PE 100 240 VAC 5 VDC PE PE 128...
Page 129: ...1 U 2 3 2 4 1 U 2 3 4 2 5 1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 129...
Page 132: ...4 3 4 5 4 2 6 1 7 ORBISPHERE 10 32505 mm Lemo 10 P8 4 132 J7 133 4 5 8 8 132...
Page 134: ...NO NC NC NO J14 1 J18 2 J19 3 3 5 10 15 20 15 2 Hach LED 10 1 134...
Page 135: ...menu Up Main Close Help Enter 4 1 Configuration Security ID 135...
Page 143: ...1 2 3 1 2 3 143...
Page 144: ...1 2 3 4 144...
Page 185: ......