the oxygen level for LDO sensor). This display can be configured to suit individual conditions and
convenience.
The K-M1100
low range
sensors measure dissolved oxygen up to a maximum value of 5000 ppb.
Below 2000 ppb the measurement cycle interval is 4 seconds for 2 channels and 6 seconds for
3 channels. Between 2000 and 3000 ppb the measurement cycle interval is 30 seconds. Above
3000 ppb the measurement cycle interval is 60 seconds. The K-M1100
high range
sensors
measure dissolved oxygen up to a maximum value of 40 ppm, the measurement cycle interval is
4 seconds for 2 channels and 6 seconds for 3 channels. Should the measured concentration go
above the maximum value for the sensor then the measurement cycle is increased to 60 seconds
and an
Out of range
message will be displayed. An arrow symbol to the right indicates if the value is
increasing, decreasing, or remaining constant. Once the measured value falls below the maximum
value, the measurement cycle returns to the pre-defined interval.
Numeric view configuration
1.
Select
Configure
from the
View
menu followed by
Conf. numeric view
to customize the display:
Option
Description
Display temperature
Select
Channel x
to display the sample temperature for that channel.
Display channel 1, 2, 3
Select yes or no
Display mini graph
Check the box to display the graph.
Display time base
Check the box to display the time base.
Upper bound
Adjust graph upper limit.
Lower bound
Adjust graph lower limit.
Time base
Adjust graph time span.
Grid button
Set up the graph to display the x or y axes, grid, or alarm thresholds.
Auto scale update button
Automatically set the graph upper and lower bounds to best fit the actual values
displayed.
Clean button
Clear the graph being displayed and restart.
Statistic view
This feature offers statistical data to match with Total Quality Management tools to better analyze
how a process behaves. The statistics are calculated from the data in the measurement file and
values updated each time a new measurement is added.
Diagnostic view
The diagnostic view contains important information but is only really useful for troubleshooting
purposes.
Measurement menu
Instrument configuration
1.
Select
Config. instrument
from the
Measurement
menu:
Option
Description
Measurement mode Continuous
mode for on line process. Default:
Continuous mode
.
Sample
mode for lab sample analysis of small individual samples, such as cans or
bottles.
English
15
Summary of Contents for 51 Series
Page 2: ...English 3 Deutsch 22 Fran ais 43 Espa ol 63 e tina 84 Svenska 104 124 Magyar 145 T rk e 165 2...
Page 127: ...ESD 127...
Page 128: ...2000 m 6562 ft 2000 m GFCI GFI PE 100 240 VAC 5 VDC PE PE 128...
Page 129: ...1 U 2 3 2 4 1 U 2 3 4 2 5 1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 129...
Page 132: ...4 3 4 5 4 2 6 1 7 ORBISPHERE 10 32505 mm Lemo 10 P8 4 132 J7 133 4 5 8 8 132...
Page 134: ...NO NC NC NO J14 1 J18 2 J19 3 3 5 10 15 20 15 2 Hach LED 10 1 134...
Page 135: ...menu Up Main Close Help Enter 4 1 Configuration Security ID 135...
Page 143: ...1 2 3 1 2 3 143...
Page 144: ...1 2 3 4 144...
Page 185: ......