1.
From the
Main
menu, select
Calibration
followed by
Gas sensor
and then
Configuration
. Make
sure the parameters are set as follows:
Option
Description
Auto-calibration
Not available for this sensor.
Manual-calibration
Make sure that the
Auto-End
box is checked.
Hold during calibration or
verification
Make sure this box is checked.
Zero calibration bottle
Make sure this is disabled by unchecking the box as it is not relevant
for this sensor.
2.
Exit from the configuration screen by pressing
OK
.
3.
Select
Calibration
and perform a manual zero calibration as described in
Zero calibration
on page 19. For high range sensors perform an additional 100% humid air calibration as
described in
100% humid air calibration (high range sensors only)
on page 19.
Manual calibration
Manual calibrations can be made at any time by following these steps:
1.
Remove the sensor from the sample line.
2.
Rinse the sensor head with clean water.
3.
Wipe the sensor head with a clean soft tissue to remove any excess moisture.
4.
If using the supplied calibration device insert the sensor into the sensor holder on top of the
calibration device. If not using the calibration device insert the sensor into the flow chamber.
5.
Flow the calibration sample through the calibration device or flow chamber as applicable. If using
the calibration device fully open the valve on the pressure reducer to give a gas flow rate of
0.1 L/min. If you are not using the supplied calibration device with pressure reducer the maximum
allowable inlet pressure must be no more than 2 bar absolute.
Note: The manufacturer recommends to keep the flow chamber at atmospheric pressure. Adjust the sample
flow before it goes into the flow chamber.
6.
Configure the calibration as described in
Calibration configuration
on page 18.
7.
Start the calibration as described in
Zero calibration
on page 19,
100% humid air calibration
(high range sensors only)
on page 19 or
High level adjustment
on page 20 depending on the
preferred calibration method.
Calibration configuration
Note: This option can also be invoked by pressing the Modify button in either the Zero calibration or High level
adjustment calibration screens.
1.
From the
Main
menu, select
Calibration
followed by
Gas sensor
and then
Configuration
.
Option
Description
Auto-calibration
Not available for this sensor.
Manual-calibration
When
Auto-End
is enabled a manual calibration will complete automatically when the
parameters defined in
Stop parameters
are reached. Press on
Configure
to set the
manual calibration parameters. If the calibration fails, the previous calibration
parameters remain unchanged and a warning message displayed.
Hold during
calibration or
verification
If checked this keeps the last measured value and stops updating the outputs during
the calibration or verification process. This avoids sending invalid information to any
connected device. At the end of a calibration, this hold remains on for a further
10 minutes to allow the system to stabilize.
18
English
Summary of Contents for 51 Series
Page 2: ...English 3 Deutsch 22 Fran ais 43 Espa ol 63 e tina 84 Svenska 104 124 Magyar 145 T rk e 165 2...
Page 127: ...ESD 127...
Page 128: ...2000 m 6562 ft 2000 m GFCI GFI PE 100 240 VAC 5 VDC PE PE 128...
Page 129: ...1 U 2 3 2 4 1 U 2 3 4 2 5 1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 129...
Page 132: ...4 3 4 5 4 2 6 1 7 ORBISPHERE 10 32505 mm Lemo 10 P8 4 132 J7 133 4 5 8 8 132...
Page 134: ...NO NC NC NO J14 1 J18 2 J19 3 3 5 10 15 20 15 2 Hach LED 10 1 134...
Page 135: ...menu Up Main Close Help Enter 4 1 Configuration Security ID 135...
Page 143: ...1 2 3 1 2 3 143...
Page 144: ...1 2 3 4 144...
Page 185: ......