SECTION 8, 

continued

8.3

Diagnostic Functions

The diagnostic mode permits access to system function information that is useful 
primarily when the instrument function is in doubt.  Hach service technicians use 
the information for precise troubleshooting, speeding repairs, and avoiding 
unnecessary service returns.

Access diagnostic information by pressing and holding the

 RIGHT ARROW

 key for 

3 seconds.  Use the  

 ARROW 

keys to edit the display to read the diagnostic code 

number of interest.  Press the 

ENTER

 key to display the diagnostic value.

8.3.1

Basic Diagnostic Codes

Access thTable 6 diagnostic information by entering the appropriate code:

Table 6 Diagnostic Codes

Code

Display

Description

00

bP on/bP of

 Keyboard Beeper On/Off

01

FS Pr/SL Pr

Fast/Slow Print Device

21

Pr ln

Printer Test

22

*

Display Test

23

*

Keyboard Test

24

*

Memory Test

25

xxxxx

90° Detector mV, Gain 1

26

xxxxx

90° Detector mV, Gain 10

27

xxxxx

90° Detector mV, Gain 100

28

xxxxx

90° Detector mV, Gain 1000

29

xxxxx

Forward Scatter Detector mV, Gain 1

30

xxxxx

Forward Scatter Detector mV, Gain 10

31

xxxxx

Forward Scatter Detector mV, Gain 100

32

xxxxx

Forward Scatter Detector mV, Gain 1000

33

xxxxx

Transmitted Detector mV, Gain 1

34

xxxxx

Transmitted Detector mV, Gain 10

35

xxxxx

Transmitted Detector mV, Gain 100

36

xxxxx

Transmitted Detector mV, Gain 1000

37

xxxxx

Transmitted  Detector mV, Gain 10,000

38

xxxxx

Transmitted Detector mV, Gain 100.000

39

xxxxx

Transmitted Detector mV, Gain 1,000,000

41

xxxxx

A/D Reference Low mV, Gain 1

45

xxxxx

A/D Reference Low mV, Gain 10

46

xxxxx

A/D Reference Low mV, Gain 100

47

xxxxx

A/D Reference Low mV, Gain 1000

48

xxxxx

A/D Reference Medium mV, Gain 1

49

xxxxx

A/D Reference Medium mV, Gain 10

50

xxxxx

A/D Reference Medium mV, Gain 100

51

xxxxx

A/D Reference Medium mV, Gain 1,000

52

xxxxx

A/D Reference High mV, Gain 1

53

xxxxx

A/D Reference High mV, Gain 10

Tài liệu sưu tầm  của Trung Sơn - www.tschem.com.vn - 028.3811.9991 - [email protected]

Tài liệu sưu tầm  của Trung Sơn - www.tschem.com.vn - 028.3811.9991 - [email protected]

Summary of Contents for 2100N

Page 1: ...oftware Version 1 Hach Company 1993 1996 1999 All rights reserved Printed in the U S A hm dh 6 99 2ed Rev 1 10 99 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i...

Page 2: ...ManVer MolyVer Mug O Meter NetSketcher NitraVer NitriVer NTrak OASIS On Site Analysis Results You Can Trust OptiQuant OriFlow OxyVer PathoScreen PbEx PermaChem PhosVer Pocket Colorimeter Pocket Pal Po...

Page 3: ...Oil Procedure 23 2 6 1 Applying Silicone Oil 23 2 6 2 Preparing Dilution Water 24 2 6 2 1 Dilution Water Filtration Procedure 24 2 6 3 Indexing and Matching Sample Cells 25 2 6 3 1 Indexing a Single S...

Page 4: ...Water 44 3 3 2 Editing Calibration Data Points 44 3 3 3 Preparing Formazin Dilutions User Selected 44 3 3 4 Calibrating the 2100N User Selected Standards 45 INSTALLATION MAINTENANCE 47 SECTION 4 AIR...

Page 5: ...69 8 3 Diagnostic Functions 70 8 3 1 Basic Diagnostic Codes 70 8 3 2 Other Instrument Diagnostics 71 8 3 2 1 Display Segments and Icons 71 8 3 2 2 Cold Start 71 8 3 2 3 Flashing 9s 71 8 3 2 4 Flashin...

Page 6: ...6 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 7: ...01 3 Radiated RF Electro Magnetic Fields ENV 50204 Radiated Electro Magnetic Field from Digital Telephones EN 61000 4 4 IEC 1000 4 4 IEC 801 4 Electrical Fast Transients Burst EN 61000 4 5 IEC 1000 4...

Page 8: ...pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment g...

Page 9: ...Air Purge Connection on page 49 Section 5 3 High Pressure Flow Cell Kit on page 58 Section 6 1 RS232 Connection on page 63 Section 7 2 Lamp Replacement on page 67 T i li u s u t m c a Trung S n www ts...

Page 10: ...10 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 11: ...0 01 NTU from 0 40 NTU with Ratio OFF Reference Conditions 0 to 40 C 0 to 90 RH Noncondensing 25 C 115 230 Vac 17 50 60 Hz Resolution 0 001 on lowest range Repeatability 1 of reading or 0 01 NTU which...

Page 12: ...0 psig 0 hose barb connection for 1 8 tubing Max 138 kPa 20 psig Dry nitrogen or instrument grade air ANSI MC 11 1 1975 Power Requirement 115 230 Vac 17 50 60 Hz 60 VA Max Automatic Power Selection Se...

Page 13: ...procedimientos de seguridad antes de manipular productos qu micos GEFAHR Das Arbeiten mit chemischen Proben Standards und Reagenzien ist mit Gefahren verbunden Es wird dem Benutzer dieser Produkte emp...

Page 14: ...14 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 15: ...th a dust cover and an instrument manual 1 3 Principle of Operation The Model 2100N Laboratory Turbidimeter is a Nephelometer with the capability to measure with either the Ratio on or Ratio off The i...

Page 16: ...N Laboratory Turbidimeterr Figure 2 2100N Laboratory Turbidimeter Optics T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com...

Page 17: ...he turbidimeter in a clean dust free environment on a bench or table that is free of vibration and that provides good air circulation around the instrument Keep the areas in the back and underneath of...

Page 18: ...18 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 19: ...tions NPDWR or National Pollutant Discharge Elimination System NPDES permits 2 2 1 Installing the Filter Assembly 1 Ensure the filter is clean and free from visible damage Note Handle the filter assem...

Page 20: ...te sur la compatibilit de l chantillon contr ler PELIGRO El Turbid metro de Laboratorio 2100N no est dise ado para usarse con muestras inflamables o que contengan hidrocarburos o cidos concentrados qu...

Page 21: ...e the filter is in place Place the sample cell in the instrument cell compartment and close the lid Note For immediate update of the display press ENTER Note When using the Flow Cell System the Flow C...

Page 22: ...the sample cell Measure samples immediately to prevent changes in sample characteristics due to temperature shifts and settling Avoid dilution whenever possible particles suspended in the original sa...

Page 23: ...y with little or no visible oil Applying excess oil may attract dirt and contaminate the sample compartment of the instrument Silicone Oil Procedure 1 Thoroughly clean and rinse the sample cell Note S...

Page 24: ...e particles from contaminating the glassware if it is not used immediately 2 6 2 1 Dilution Water Filtration Procedure 1 Attach the syringe to the three way valve by gently twisting the square end int...

Page 25: ...ngle Sample Cell Add an orientation mark to a single sample cell as follows 1 Fill the clean cell to the line with high quality water and cap the sample cell refer to Section 2 6 2 2 Wipe the sample c...

Page 26: ...ilicone oil Note For more information see Section 2 5 1 Cleaning Sample Cells Section 2 6 1 Applying Silicone Oil Section 2 6 2 Preparing Dilution Water 2 Make sure the filter is in place Insert the f...

Page 27: ...or vacuum degassing An electric or hand operated pump equivalent to Cat No 14697 00 or 14283 00 respectively also may be used Figure 4 Sample Degassing 2 6 4 2 Addition of Surfactant Limit the use of...

Page 28: ...crease the ultrasound waves probably have started to alter the suspended particles Note the treatment time before the turbidity increase and record it as the maximum time limit for ultrasonic treatmen...

Page 29: ...particles and little light is available for scattering This results in a negative interference the measured turbidity is lower than the actual turbidity This condition is called going blind Light abso...

Page 30: ...100 mL total 2 Calculate the Final Turbidity Value For example if the measured turbidity value is 1100 NTU the final turbidity value is calculated as Dilution Factor Total Volume Sample Volume Diluti...

Page 31: ...mm O D The 12 to 13 mm adapter accommodates either 12 mm or 13 mm tubes The minimum sample volumes that must be used are 2 5 mL for 12 mm tubes 3 5 mL for 13 mm tubes 5 mL for 16 mm tubes and 7 mL for...

Page 32: ...ese values for standardization checks on other instruments Gelex standards remain stable when cared for properly Handle with care and store them in their protective box at room temperature The Gelex s...

Page 33: ...dimeter Information on the functions of each of these features is provided in Table 2 and supplemented with additional details in Section 3 1 1 on page 35 through Section 3 1 8 on page 36 Figure 8 Mod...

Page 34: ...bration or possibly an instrument malfunction If the CAL annunciator flashes the instrument must be recalibrated 13 RANGE Selects Auto Ranging or Manual Ranging Pressing RANGE steps the instrument thr...

Page 35: ...UNITS EXIT Key This key selects NTU EBC or Nephelo measurement units If power is turned off the instrument defaults to the last unit selected when power is restored Press the UNITS EXIT key to select...

Page 36: ...buffer begins accumulation of new data To begin measurement of a standard during calibration press ENTER while in the calibration mode 3 1 8 Using the ARROW Keys The ARROW keys are used to edit the d...

Page 37: ...libration using pre calibrated Gelex Secondary Standards refer to Section 3 2 5 on page 42 If the reading in the range of use is not within 5 of the standard s assigned value recalibrate using Formazi...

Page 38: ...f dilution water to a clean 200 mL Class A volumetric flask With a TenSette Pipet add 1 00 mL of well mixed 4000 FNU formazin stock solution to the 200 mL flask Dilute to the mark with dilution water...

Page 39: ...ed and recorded simultaneously Note Calibration for EBC and NEPH units of measure is set automatically Note Upon entering the Calibration Mode Automatic Range Signal Averaging On and NTU units are aut...

Page 40: ...display counts down from 60 to 0 and displays the turbidity compensated for dilution water turbidity The instrument automatically increments to the next standard the display shows 200 0 NTU and the S2...

Page 41: ...urns the instrument to the measurement mode Note If power is lost during calibration new calibration data is lost and the old calibration remains in effect To exit calibration without saving new value...

Page 42: ...ent with the triangle on the vial aligned with the index mark on the instrument sample compartment Close the sample cell cover 5 Press the ENTER key Record the value displayed Remove the standard from...

Page 43: ...eim Umgang mit diesen Chemikalien notwendigen Vorsichtsma nahmen Risiken und Notfallschutzma nahmen vertraut machen Es wird empfohlen in allen Situationen in denen mit einem Kontakt von Chemikalien zu...

Page 44: ...de causing the decimal point to flash Use the RIGHT ARROW key to move the decimal point to the appropriate location Pressing the ENTER key accepts the new decimal location and causes the 2 to flash Be...

Page 45: ...4 Calibrating the 2100N with two exceptions Standards are values other than those used in steps 8 10 and 11 Before pressing the ENTER key to measure each standard edit the displayed value reflecting t...

Page 46: ...46 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 47: ...instrucciones a fin de evitar accidentes personales y da os al instrumento Las tareas de instalaci n y mantenimiento descritas en la presente secci n deber n ser efectuadas nicamente por personas deb...

Page 48: ...48 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 49: ...is expected to be near or below 2 C 35 F use a desiccant dryer and particle filter in the air line to ensure the dew point of the air purge is below the sample temperature see Figure 9 If only shop a...

Page 50: ...50 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 51: ...acides concentr s ou bases concentr es qui peuvent attaquer les parties au contact du liquide Effectuer des essais avant l utilisation des cuves circulation si la compatibi lit de l chantillon est do...

Page 52: ...into the top of the cell A baffle deflects the incoming sample to the side wall of the cell minimizing turbulence in the light path Sample is discharged from top and bottom outlets The top outlet col...

Page 53: ...e circulation basse pression manuelle ou automatis e est con ue pour utilisation sous faible pression uniquement 0 34 bar 34 kPa 5 psig PRECAUCION La C lula de Flujo de faja presi n tanto manual como...

Page 54: ...servoir on to the support rod 5 Slide the Collection Drain Assembly on to the support rod 6 Place the instrument right side up and install the support rod into the base 5 2 1 2 Assembling The Flow Cel...

Page 55: ...Then fill the system with water to ensure all connections are water tight before inserting the Flow Cell into the sample compartment of the instrument Once filled inspect the system for leaks Also mak...

Page 56: ...rol cable used with the Hach Model 2100AN Laboratory Turbidimeter for automated operation The cable is not used with the Model 2100N Turbidimeter Refer to Section 5 2 1 1 and Section 5 2 1 2 for assem...

Page 57: ...supply into an appropriate wall receptacle 5 2 2 2 Using the Automated Flow Cell Kit Note Assemble the Flow Cell tubing and stand Then fill the system with water to ensure all connections are water t...

Page 58: ...ubbles Do not use the system for monitoring flammable solutions solvents strong acids or strong bases Do not exceed the recommended maximum sample pressure of 34 kPa 5 psig Fill the system with distil...

Page 59: ...ession Fitting Connection Note Hand tighten the compression fitting nuts Excessive force may damage the fitting or cap assembly see Figure 17 Figure 16 illustrates a suggested hookup using a pressure...

Page 60: ...e 14 Make sure the O ring is in place in the cap assembly to ensure a good seal hand tighten Wipe the glass surface free of fingerprints and smudges Apply a thin even film of silicone oil to the outsi...

Page 61: ...d the unit by the glass cell Use a protective shield between the operator and the cell DANGER Porter des lunettes de protection et manipuler avec une extr me prudence lors de la pressurisation du syst...

Page 62: ...ng Coat the outside surface of the glass Flow Cell parts with a thin film of silicon oil before installation in the instrument see Section 2 6 1 on page 23 Note Always test the system for leaks before...

Page 63: ...ailable as an optional accessory refer to on page 74 With the use of a serial to parallel converter the data string transmitted from the 2100N Turbidimeter prints on any Epson compatible parallel prin...

Page 64: ...put to drive a printer Figure 22 provides a sample printout from the forty column Citizen printer listed in Optional Accessories on page 74 Instrument DB9 Female Printer DB25 Male 3 3 5 7 6 20 SHELL 1...

Page 65: ...iDP 562RSL II printer board as follows see Figure 23 and Table 4 1 Set switches 6 7 9 10 to ON 2 Set switches 1 2 3 4 5 8 to OFF 3 Refer to Setting of Preset Jumper in the Citizen Printer Manual Figu...

Page 66: ...ogram to 1200 baud 8 data bits no parity 1 stop bit The following RS232 command set is available when a computer is connected to the 2100N Key in VAL for value and press enter on the computer keyboard...

Page 67: ...cement Use only the Lamp Replacement Kit Cat No 47089 00 The lamp assembly includes the lamp with leads the lamp retainer and the lamp holder Replace the lamp as follows 1 Turn the instrument off and...

Page 68: ...cause damage Clean the lamp with alcohol if necessary 5 Mount the new lamp assembly in the instrument 6 Insert the lamp leads under the clamping plates in the lamp terminal block Lamp leads are not p...

Page 69: ...n standards have the same value or their difference is less than 60 0 NTU Standard 1 is too low 10 NTU Recheck preparation of standards and repeat calibration Err03 Low light error Reinsert sample Che...

Page 70: ...90 Detector mV Gain 100 28 xxxxx 90 Detector mV Gain 1000 29 xxxxx Forward Scatter Detector mV Gain 1 30 xxxxx Forward Scatter Detector mV Gain 10 31 xxxxx Forward Scatter Detector mV Gain 100 32 xxx...

Page 71: ...ample is 40 NTU 268 Nephelos or 9 8 EBCs with the Ratio off refer to Section 2 8 on page 29 8 3 2 4 Flashing 0s If the display flashes 0s the measurement indicates a negative turbidity value Recalibra...

Page 72: ...72 T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 73: ...make With that in mind we have compiled the following information for your convenience 73 GENERAL INFORMATION T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u...

Page 74: ...ultrasonic 24895 00 Cable computer 9 pin to 9 pin 49502 00 Cable for Citizen printer Cat No 25933 00 and 25933 02 9 pin to 25 pin 49503 00 Calibration Kit StablCal for 2100N Turbidimeter 0 1 20 200 1...

Page 75: ...acuum hand operated 14283 00 Pump vacuum pressure portable 120 V 60 Hz 1 3 cfm 14697 00 Pump vacuum pressure portable 240 V 50 Hz 1 3 cfm 14697 02 Ribbon Cartridge iDP 562 for Citizen Printer Model iD...

Page 76: ...76 HOW TO ORDER T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn T i li u s u t m c a Trung S n www tschem com vn 028 3811 9991 info tschem com vn...

Page 77: ...merica the Caribbean the Far East the Indian Subcontinent Africa Europe or the Middle East Hach Company World Headquarters P O Box 389 Loveland Colorado 80539 0389 U S A Telephone 970 669 3050 FAX 970...

Page 78: ...THE PRECEDING SENTENCE HACH MAKES NO WARRANTY OF ANY KIND WHATSOEVER WITH RESPECT TO ANY PRODUCTS HACH EXPRESSLY DISCLAIMS ANY WARRANTIES IMPLIED BY LAW INCLUDING BUT NOT BINDING TO ANY WARRANTY OF M...

Page 79: ...and held radio communication device at least 3 to 4 meters from the instrument to provide isolation As shown below the instrument reading may deviate slightly win the presence of specific frequencies...

Page 80: ...TION AND ORDERING Call 800 227 4224 Contact the HACH office or distributor serving you www hach com techhelp hach com In the U S A Outside the U S A On the Worldwide Web E mail toll free T i li u s u...

Reviews: