SPECIFICATIONS,
continued
Sample Cells: 95 mm high x 25 mm diam. (3.74 in high x 1 in diameter).
Borosilicate glass with rubber-lined screw caps.
Sample Required: 30 mL (1 oz.) minimum
Secondary Standards: Gelex® Secondary Standards
Temperature:
Storage Temperature: -40 to 60 °C (-40 to 140 °F)
Operating Temperature: 0 to 40 °C (32 to 104 °F)
Sample Temperature: 0 to 95 °C
Operating Humidity Range: 0 to 90% RH Noncondensing @ 25 °C; 0 to 75%
RH Noncondensing @ 40 °C
Instrument Stabilization Time: 30 min. with ratio on, 60 min. with ratio off;
typical application leaves instrument on 24 hrs/day
Air Purge: 0.1 scfm at 69 kPa (10 psig: 0 hose barb connection for 1/8ð tubing,
Max 138 kPa (20 psig). Dry nitrogen or instrument grade air
(ANSI MC 11.1, 1975)
Power Requirement: 115/230 Vac ±17%, 50/60 Hz, 60 VA Max (Automatic
Power Selection)
Serial I/O: RS232C serial interface via DB9 subminiature D shell connector for
data output to computer or printer, and data input (command). No handshaking.
Factory set for a 1200 baud rate, one stop bit, no parity, eight bit character length.
Enclosure: High-impact polycarbonate plastic
Dimensions: 30.5 x 40 x 15.6 cm (12 x 153/4 x 61/8 in.)
Instrument Weight: 3.43 kg (7 lbs, 9 oz)
Shipping Weight (with standard accessories): 5.58 kg (12 lbs. 5 oz.)