3
Section 1
Specifications
Specifications are subject to change without notice.
General
Size
170.94 mm x 58.42 mm x 165.10 mm (6.7" W x 4.5" H x 11.7" D)
Weight
3.0 kg (6.8 lb)
Maximum count displayed
9,999,999
Sample flow rate
0.1 cfm
Number of size ranges
2, 4, 5, 6 (specified at the time of order)
Particle size ranges
MET ONE 237A—0.5 and 5.0 micron (2 channel); 0.5, 0.7, 1.0 and 5.0 micron (4
channel); 0.5, 0.7, 1.0, 2.0 and 5.0 micron (5 channel); 0.5, 0.7, 1.0, 2.0, 3.0 and
5.0 micron (6 channel)
MET ONE 237B—0.3 and 5.0 micron (2 channel); 0.3, 0.5, 1.0 and 5.0 micron (4
channel); 0.3, 0.5, 0.7, 1.0 and 5.0 micron (5 channel); 0.3, 0.5, 0.7, 1.0, 2.0 and
5.0 micron (6 channel)
Coincidence error
Less than 5% at 2,000,000 particles/cubic feet
Power
+5 W, +10 V (approximately) (+6 V with battery pack)
Light source
Laser diode
Sample/hold times
1 second to 24 hours
Count alarms
1 to 9,999,999 counts
Data storage
500 samples, rotating buffer
Count cycles
Up to 100 in Auto mode
Locations
Up to 250 (number appears on the printout)
Output
RS232/RS485 for computer
Pump type
Carbon-vane, 0.1 cfm
Battery type
Rechargeable NiCd, 3500 mAH
Operation time
4 hours printing, 8 hours without printing
Environment
Operating temperature
2 to 29 °C (55 to 84 °F)
Operating humidity
10 to 85% relative, non-condensing
Storage temperature
-40 to 71 °C (–40 to 160 °F)
Storage humidity
Up to 98% relative, non-condensing
Summary of Contents for MET ONE 237AB
Page 1: ...701076 MET ONE 237AB Particle Detector USER MANUAL June 2008 Edition 5 ...
Page 3: ......
Page 7: ...4 Specifications ...
Page 17: ...14 System Start Up ...
Page 31: ...28 Maintenance ...
Page 37: ...34 Replacement Parts and Accessories ...
Page 38: ...36 ...
Page 40: ...38 ...
Page 48: ...46 ...
Page 54: ...52 Index ...