Troubleshooting Procedures
7.3 Pump troubleshooting
If the vacuum pump does not operate, or exhibits any questionable symptoms such as
unusual sound or vibration, it may be removed for maintenance to restore it to optimum
operation.
7.3.1 Disassembly and cleaning of the pump
Prerequisites:
•
1.27 mm or 1.5 mm Allen wrench
To disassemble and clean the pump:
1.
Open the counter as described in
, step 1 to step 3.
2.
Remove the pump from the chassis clip.
3.
Remove the pump electrical connection.
4.
Disconnect the tubing from the pump.
Note:
Make sure to note the attachments to the Vacuum (V) fitting and the Pressure (P) fitting.
5.
Use a 1.27 mm or 1.5 mm hex key (Allen) wrench to loosen the set screw on the
bracket on top of the pump (refer to
).
6.
Push the bracket to one side and remove it from the pump (refer to
).
7.
Remove the top of the pump. Note the locator pin (refer to
8.
Remove pump body (refer to
).
9.
Remove the carbon vanes and make sure the rotor turns freely (refer to
).
10.
Perform steps 1 to 9 in the reverse order to assemble the pump.
Figure 12 Loosening of the set screw
1
Hex Allen key wrench