32
Troubleshooting Procedures
7.3.2 Pump optimization
To optimize the pump:
1.
Assemble the pump again. Loosen the setscrew plug in top bracket and make the
electrical connection to the MET ONE 237 particle counter.
2.
Turn on the power to the MET ONE 237 particle counter to initiate a count cycle and
turn on the pump.
3.
Hold the pump motor steady.
4.
Rotate the entire pump head in a 180° arc (refer to
).
5.
As the pump head is rotated listen for the point where the Rotations per minute
(RPM) are the greatest.
6.
Keep the pump head steady and tighten the set screw.
Figure 16 Removal of carbon vanes
1
Carbon vane
Figure 17 Rotation of the pump head