22
Operation
.
To program the MET ONE 237 particle counter for Concentration mode:
1.
Press and hold the
MODE
key until
CONCEN
is lit.
2.
Press and hold the
SELECT
key
until
OPT
is displayed.
3.
Press the
UP
or
DOWN
arrow until the 3-digit display shows VOL (volume). The 7-digit
display will show either L (liter), 1000 L or CF (cubic foot). L is the default setting from
the factory and the concentration estimates will be expressed as counts per liter.
To change the volume settings:
1.
Press and hold the
PROG
key so the light on the
PROG
key turns red and the volume
units in the 7-digit display begins to flash.
2.
Press the
UP
or
DOWN
arrow till the required setting is displayed.
3.
Press and hold the
PROG
key to save the settings. The light on the
PROG
key will
turn off and the display will not flash now.
5.3.3.1 Set the sample period
To set the sample PERIOD to the desired length:
1.
Press and hold the
SELECT
key until the PER light is lit. The 7-digit display shows the
current period setting in HH:MM:SS format, for example:
00:00:06
0.
Figure 9 Calculation using 3-second period
1
3-second period
3
Recalculation and display update
2
Seconds
Figure 10 Calculation using 10-second period
1
10-second period
3
Recalculation and display update
2
Seconds
Summary of Contents for MET ONE 237AB
Page 1: ...701076 MET ONE 237AB Particle Detector USER MANUAL June 2008 Edition 5 ...
Page 3: ......
Page 7: ...4 Specifications ...
Page 17: ...14 System Start Up ...
Page 31: ...28 Maintenance ...
Page 37: ...34 Replacement Parts and Accessories ...
Page 38: ...36 ...
Page 40: ...38 ...
Page 48: ...46 ...
Page 54: ...52 Index ...