4.
Push the arrow keys to scroll through the application parameters. The
Edit
key is only available
when a parameter can be changed.
Note: More parameters can be changed with the instrument in expert mode.
5.
Push
Edit
to change the parameter. Enter the new details or select from a list.
Install the applications
Use the supplied USB key to install the applications. The instrument can install a maximum of five
applications. The installed applications are shown on the top line of the display. If any errors occur
during installation, refer to
on page 22.
1.
Push
Home
to go to the main menu.
2.
Connect the USB key to the USB port on the side of the instrument. The applications on the USB
key show on the display.
3.
Push the arrow keys to highlight and select an application to install. Push the left or right arrow
key to select it. Do this step again to select additional applications to install.
4.
Push
Import
to install the selected applications.
5.
Push
OK
to complete the installation. The installed applications show on the main menu.
Note: To install more applications, push Home to go to the main menu, then remove the USB key and
reconnect it.
Prepare the instrument for measurement
1.
From the main menu, select
Purge
, then push
Start
. All attached devices are listed.
2.
Select
All elements
to purge all the attached devices, or select one device to purge. Push
Select
. Air is removed from the device and filled with liquid from the bottle.
3.
Push
OK
when the operation has completed.
4.
Make sure that there are no air bubbles in the device. Do step
bubbles.
5.
Select the next device to purge if individual devices are being selected.
6.
Push
Exit
when all the tubes are filled with reagent and the device has no air bubbles.
Note: If a few small air bubbles can be seen on the inner wall and/or piston of the syringe, they can be left
without effecting system performance.
Standard operations
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
C A U T I O N
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
C A U T I O N
Personal injury hazard. Never use the instrument without the syringe protection cover in place.
C A U T I O N
Chemical exposure hazard. Never remove the stir bar from the beaker before the end of a titration.
16
English
Summary of Contents for TitraLab KF1000 Series
Page 21: ...English 21 ...
Page 130: ...130 Português ...
Page 151: ...Čeština 151 ...
Page 172: ...172 Dansk ...
Page 237: ...Svenska 237 ...
Page 322: ...322 Română ...
Page 343: ...Türkçe 343 ...
Page 364: ...364 Slovenský jazyk ...
Page 385: ...Slovenski 385 ...
Page 406: ...406 Hrvatski ...
Page 428: ...428 Ελληνικά ...
Page 431: ......