4.11. Data Logger
The si629 C is able to store up to 200 data in memory. When the data logger is full, data will be stored
replacing the eldest ones. Data can be stored following two different criteria:
Interval:
data are stored with the frequency programmed in
Interval:
XX min.
Event:
data are stored only if the values are higher o lower than the
programmed value, and with the selected frequency.
It displays all stored data. The first record will be
displayed, and then it will automatically go on
one by one every 1 sec.
Press
to quit the View display.
Press
to stop the automatic view. The display
will be frozen at the last displayed record.
Press to see data in increasing order.
Press to see data in decreasing order.
DATA LOGGER
CODE 000
MEASURING
22.3°C
545 µS/cm
@
25°
D. LOG.
> View
Program
Erase
MEASURE
MEASURING 22.3°C
545 µS/cm
@
25°
Press
to select the desired option.
Select START to start the storage
of data.
Select STOP to end this option.
To empty the
data logger
select the option
erase.
VIEW
> Max.
Min.
All
Min 12-04 / 11:21
520 µS/cm 25.2°C
Event
EC <XX.XX
t= 30 m
Max 12-04 / 12:22
1250 µS/cm 25.2°C
Intervalo
t= 30 m
001 12-04 / 11:01
545 µS/cm 25.2°C
Date
10-03-2004
PROG.
> Interval *
Event
Start
Stop *
Date
Hoor
Hour
12:04:56
ERASE
DATA LOGGER
ERASE
Press
to select the
desired option.
Select the value and the unit (sec., min. or hours)
It displays only the maximum stored reading.
It displays only the minimum stored reading.
4. Operation
Press
until the desired option is reached.
Enter code 100 and y
accept with
si629 C_HACH LANGE_Ed 0510.indd ANG:25
si629 C_HACH LANGE_Ed 0510.indd ANG:25
20/5/10 12:06:58
20/5/10 12:06:58