Product Overview
The LINK2SC software function enables optimum interaction of the
process measuring equipment and the photometer in the laboratory. It
enables direct measurement correction of the online sc probe via a
photometer operating under laboratory conditions (refer to
Introduction to matrix correction
). LINK2SC also enables laboratory
control measurement. Measurement data is transferred from the
sc probe to the photometer, where it is then archived together with the
photometric reference data.
One or more online measurement values are used to create a
job
on the
sc controller; this job is transferred to the photometer via a local network
connection or via SD memory card and USB adapter. A job is an XML
file containing measurement values and additional controller and probe
information.
The water probe drawn at the same time as this is analyzed in the
laboratory in line with the same parameters.
In the photometer, the photometric measured value is assigned to the
probe value in the
job
. If the job is then sent back to the sc controller,
the probe is corrected automatically.
Data transfer is either via SD memory card or local network (LAN)
1
.
Figure 1 Communication between the process measurement
equipment and the photometer in the laboratory
Introduction to matrix correction
There are various options available for sc probes to correct the probe
value by means of laboratory values (as a reference value).
This example uses the AN-ISE sc probe, for which the sample
laboratory value can be entered as either nitrate nitrogen (NO
3
-N) or
ammonium nitrogen (NH
4
-N). This laboratory value corrects the value
measured by the probe. For precise information, please refer to the
calibration/matrix correction chapter of the AN-ISE sc user manual.
LINK2SC supports matrix correction 1 and 2 and makes an independent
decision as to which of the two correction types is to be used for the
application. The comparison measurements should be taken on different
days and at different times in order to record a maximum number of
changes in concentration. The following table describes the two
correction types in more detail.
Product
Software version
sc1000 controller
V 1.3
sc200 controller
V 1.0
DR 3900
V 1.0
AN-ISE sc
V 1.03
1. Only valid for sc1000 controllers.
1
Water probe drawn at the same
time
4
sc controller
2
Photometer and barcode test
5
sc probe
3
Data transfer: SD memory card
or LAN
1
6
Process basin
Summary of Contents for DR 3900
Page 10: ...Български 10 ...
Page 18: ...Čeština 18 ...
Page 26: ...Dansk 26 ...
Page 34: ...Deutsch 34 ...
Page 42: ...Ελληνικά 42 ...
Page 50: ...English 50 ...
Page 58: ...Español 58 ...
Page 66: ...Suomi 66 ...
Page 74: ...Français 74 ...
Page 82: ...Hrvatski 82 ...
Page 90: ...Magyar 90 ...
Page 98: ...Italiano 98 ...
Page 106: ...Nederlands 106 ...
Page 114: ...Polski 114 ...
Page 130: ...Română 130 ...
Page 138: ...Slovenský jazyk 138 ...
Page 146: ...Slovenščina 146 ...
Page 154: ...Svenska 154 ...
Page 162: ...Türkçe 162 ...
Page 163: ...www hach com www hach lange com HACH LANGE GmbH 2011 All rights reserved Printed in Germany ...